    NI 5412 Specifications NI PXI/PCI-5412 14-Bit 100 MS/s Arbitrary Waveform Generator Unless otherwise noted, the followi ng conditions were used for each specification: • Interpola tion set to maximum allo wed factor f or a gi v en sampl e rate. • Signals terminated with 50 Ω . • Lo w-Gain Amplif ie r Path set to 2 V pk-pk , and High-Gain Am ...

  • National Instruments NI 5412 - page 2

    NI 5412 Specifications 2 ni.com CH 0 (Channel 0 Analog Output , Front Panel Connector) Table 1. Specif ication Va l u e Comments Number of Channels 1 — Connector SMB (jack) — Output Voltage Characteristics Output Paths The software-selectable Main Ou tput Path setting provides full-scale vo ltages from 12.00 V pk-pk to 5.64 mV pk-pk into a 50 ? ...

  • National Instruments NI 5412 - page 3

    © National Instruments Corpor ation 3 NI 5412 Specifications Maximum Output Voltage Maximum Output Vo l t a g e Pat h Load Maximum Output V oltage (V pk -pk ) The Maxi mum Output V oltage of the NI 5412 is determined by the Amplitude Range and the Of fset Range. Low- Gain Amplif ier 50 Ω ±1.000 1 k Ω ±1.905 Open ±2.000 High- Gain Amplif ier ...

  • National Instruments NI 5412 - page 4

    NI 5412 Specifications 4 ni.com Frequency and Transient Response Bandwidth 20 MHz –3 dB Digital Interpolation Filter Software-selectable Finite Im pulse Response (FIR) filter . A vailable interpolation f actors are 2, 4, or 8. — Passband Flatness Low-Gain and High-Gain Amplifiers Path — ±1.0 dB from DC to 6 MHz Pulse Response Pat h All value ...

  • National Instruments NI 5412 - page 5

    © National Instruments Corpor ation 5 NI 5412 Specifications Figure 1. Pulse Response, Low-Gai n Amplifier Path 50 Ω Load –2.0 –1.6 –0.4 –1.2 0.0 –0.8 0.4 0.8 1.2 1.6 2.0 0 0.2 0.4 0.6 0.8 1.0 Time ( µ s ) Amplitude (V) ...

  • National Instruments NI 5412 - page 6

    NI 5412 Specifications 6 ni.com Table 1. (Continued) Specification Va l u e Comments Suggested Maximum Frequencies for Common F unctions Function Path — Lo w-Gain Amplif ier High-Gain Amplifier Sine 20 MHz 20 MHz Square 5 MHz 5 MHz Ramp 1 MHz 1 MHz T r iangle 1 MHz 1 MHz Spectral Characteristics Spurious-Free Dynamic Range (SFDR) without Harmonic ...

  • National Instruments NI 5412 - page 7

    © National Instruments Corpor ation 7 NI 5412 Specifications Sample Clock Spectral Characteristics (Continued) A verage Noise Density Pat h Amplitude Range A verage Noise Density A verage Noise Density at small amplitudes is limited by a –148 dBm/Hz noise floor . V pk-pk dBm dBm/Hz dBFS/ Hz Lo w-Gain 2 10 45 –134 –144 High- Gain 12 25.6 251 ...

  • National Instruments NI 5412 - page 8

    NI 5412 Specifications 8 ni.com Sample Rate Rang e and Resolution Sample Clock Source Sample Rate Range Sample Rate Resolution — Divide-b y- N 23.8 4 S/s to 100 MS/s Settable to ( 100 MS/s) / N (1 ≤ N ≤ 4,194,304) High- Resolution 10 S/s to 100 MS/s 1.06 µHz CLK IN 200 kS/s to 1 05 MS/s Resolution determined by external clock source. Externa ...

  • National Instruments NI 5412 - page 9

    © National Instruments Corpor ation 9 NI 5412 Specifications System Phase Noise and Jitter (10 MHz Carrier) Sample Clock Source System Phase Noise Density (dBc/Hz) Offset System Output Jitter (Integrated from 100 Hz to 100 kHz) 1. High- Resolution specifications va r y wi t h Sample Rate. 2. All values are typical. 100 Hz 1 kHz 10 kHz NI PXI-5412 ...

  • National Instruments NI 5412 - page 10

    NI 5412 Specification s 10 ni.com Onboard Clock (Internal VCXO) Phase-Locked Loop (PLL) Reference Clock Table 3. Specif ication Va l u e Comments Clock Source Internal sample clocks can either be locked to a Reference Clock using a phase-locked loop or be deriv ed from th e onboard VCXO frequency reference. — Frequenc y Accuracy ±25 ppm — Tabl ...

  • National Instruments NI 5412 - page 11

    © National Instruments Corpor ation 11 NI 5412 Specifications CLK IN (Sample Clock and Reference Cloc k Input, Front Panel Connector) Table 5. Specif ication Va l u e Comments Connector SMB (jack) — Direction Input — Destinations 1. Sample Clock 2. PLL Reference Clock — Frequenc y Range 1 MHz to 105 MHz (Sample Clock destination and sine wa ...

  • National Instruments NI 5412 - page 12

    NI 5412 Specification s 12 ni.com PFI 0 and PFI 1 (Programmable Function Interfac e, Front Panel Connectors) Table 6. Specif ication Va l u e Comments Connectors T wo SMB (jack) — Direction Bi-directional — Frequenc y Range DC to 105 MHz — As an Input (Trigger) Destinations Start T rigger — Maximum Input Overload –2 V to +7 V — V IH 2.0 ...

  • National Instruments NI 5412 - page 13

    © National Instruments Corpor ation 13 NI 5412 Specifications Start T rigger V OH Minimum: 2.9 V (open load), 1.4 V (50 Ω load) Output dri vers are +3.3 V TTL compatible. Measured with a 1 m cable. V OL Maximum: 0.2 V (open load), 0.2 V (50 Ω load) Rise/F all T ime (20% to 80%) ≤ 2.0 ns Load of 10 pF . Table 7. Specif ication Va l u e Commen ...

  • National Instruments NI 5412 - page 14

    NI 5412 Specification s 14 ni.com Delay from Start T rigger to CH 0 Analog Output Interpolation Factor T ypical Delay Refer to t s2 at NI Signal Generators Help»Devices» NI 5412» NI <bus>-5412» Triggering» Trigger Timing . Digital Inter polation F ilter disabled. 43 Sample Clock Periods + 110 ns 2 57 Sample Clock Periods + 110 ns 4 63 Sa ...

  • National Instruments NI 5412 - page 15

    © National Instruments Corpor ation 15 NI 5412 Specifications Markers Table 8. Specif ication Va l u e Comments Destinations 1. PFI<0..1> (SMB front panel connectors) 2. NI PXI-5412 —PXI_T rig<0..6> (backplane connector) NI PCI-5412 —R TSI<0..6> — Quantity One Marker per Se gment. — Quantum Marker position must be placed ...

  • National Instruments NI 5412 - page 16

    NI 5412 Specification s 16 ni.com W aveform and Instruction Memor y Utilization Table 9. Specif ication Va l u e Comments Memory Usage The NI 5412 uses the Synchronization and Memory Core (SMC) technology in which wav eforms and instructions share onboard memory . Parameters, such as number of segments in sequence list, maximum number of waveforms ...

  • National Instruments NI 5412 - page 17

    © National Instruments Corpor ation 17 NI 5412 Specifications Loop Count 1 to 16,777,215. Burst trigger: Unlimited — Quantum W aveform size must be an integer multiple of four samp les. — Memory Limits 8 MB Standard 32 MB Option 256 MB Option All trigger modes except where noted. Arbitrary W av eform Mode, Maximum W av eform Memory 4,194,176 S ...

  • National Instruments NI 5412 - page 18

    NI 5412 Specification s 18 ni.com Calibration Power Table 10. Specif ication Va l u e Comments Self-Calibrati on An onboard, 24-bit ADC and precision voltage reference are used to calibrate the DC gain and offset. The self-calibration is initiated by the user through the software and takes approximately 75 seconds to complete. — External Calibrat ...

  • National Instruments NI 5412 - page 19

    © National Instruments Corpor ation 19 NI 5412 Specifications Software Table 12. Specif ication Va l u e Comments Dri ver Software NI-FGEN 2.3 or later ve rsion. NI-FGEN is an IVI-compliant driv er that all ows you to conf igure, control , and calibrate the NI 5412. NI -FGEN prov ides application programming interfaces for man y de velopment en vi ...

  • National Instruments NI 5412 - page 20

    NI 5412 Specification s 20 ni.com Environment NI PXI-5412 Environment Note T o ensure that the NI PXI-5412 cools ef fectively , follo w the guidelines in the Maintain Forced-Air Cooling Note to Users incl uded in the NI 5412 kit. The NI PXI-5412 is intended for indoor use only . Table 13. Specifications Va l u e Comments Operating T emperature 0 º ...

  • National Instruments NI 5412 - page 21

    © National Instruments Corpor ation 21 NI 5412 Specifications NI PCI-5412 Environment Note T o ensure that the NI PCI-5412 cools ef fectively , follo w the guidelines in the Maintain Forced -Air Cooling Note to Users inclu ded in the NI 5412 kit. Also, to maxim ize airflo w and extend the life of the device, lea ve any adjacent PCI slots empty . T ...

  • National Instruments NI 5412 - page 22

    NI 5412 Specification s 22 ni.com Safety , Electromagnetic Compatibility , and CE Compliance Table 15. Specification Va l u e Comments Safety The NI 5412 meets the requirements of the following standards for safety and electrical equipment for measurement, control, and laboratory use: • IEC 61010 -1, EN 61010-1 • UL 61010-1 • CAN/CSA C22.2 No ...

  • National Instruments NI 5412 - page 23

    © National Instruments Corpor ation 23 NI 5412 Specifications EMC/EMI CE, C-T ick, and FCC Part 15 (Class A) Compliant Notes : 1. This device is not intended for , and is restricted from, use in residential areas. 2. For EMC compliance, op erate this device with shielded cabling. 3. When connected to other test objects, thi s product may cause rad ...

  • National Instruments NI 5412 - page 24

    NI 5412 Specification s 24 ni.com Physical Table 16. Specif ication Va l u e Comments Dimensions NI PXI-5412 NI PCI-5412 — Single 3U PXI slot. CompactPCI compatib le. 2.0 × 13.0 × 21.6 cm (0.8 × 5.1 × 8.5 inches) 34.07 × 10.67 × 2.03 cm (13.4 × 4.20 × 0.8 inches) We i g h t 340 g (11 oz) 480 g (17 oz) — Front Panel Connectors Label Func ...

  • National Instruments NI 5412 - page 25

    © National Instruments Corpor ation 25 NI 5412 Specifications Where to Go for Support The National Instrum ents W eb site is y our complete resource for technical support. At ni.com/support you hav e access to ev erything from troubleshooting and ap plication dev el opment self-help resources to email and phone assistance from NI Application Engin ...

  • National Instruments NI 5412 - page 26

    National Instruments, NI, ni.com, and LabVIEW are tradem arks of National Instruments Corporation. Refer to t he Terms of Use section on ni.com/legal for more information a bout National Instruments trademarks. Other product and compan y names mentioned herein are trademarks or trade names of their respective companie s . For patents covering Natio ...

メーカー National Instruments カテゴリー Portable Generator

National Instruments NI 5412のメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- NI 5412の取扱説明書
- National Instrumentsの製品カード
- パンフレット
- またはNational Instruments NI 5412の消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、National Instruments NI 5412の取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、National Instruments NI 5412の取り付け説明書、サービスマニュアル、簡易説明書、またはNational Instruments NI 5412のユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、National Instruments NI 5412の製品を使用するにあたって最も人気のある説明書を閲覧できます。

National Instruments NI 5412デバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがNational Instruments NI 5412を使用するのを助ける技術的文書のことです。説明書は通常、全てのNational Instruments NI 5412ユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

National Instrumentsの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. National Instruments NI 5412の説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははNational Instruments NI 5412の目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。National Instruments NI 5412に関してこのドキュメントで見つけることができる全てのヒントの目次
3. National Instruments NI 5412デバイスの基本機能を使うにあたってのヒント。 National Instruments NI 5412のユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。National Instruments NI 5412に関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるNational Instruments NI 5412のメーカー/サービスへの連絡先に関する情報。

National Instruments NI 5412についてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもNational Instruments NI 5412の問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがNational Instruments NI 5412で同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)