    Micro Motion ® T -Series Sensor Inst allation Manual Inst allation Manual P/N 20002172, Rev . B July 2008 ...

  • Emerson Process Management T-SERIES - page 2

    ©2008, Micro Motion, Inc. All right s reserved. ELITE and ProLink are registered tr ademarks, and MVD and MVD Direct Connect are trademarks of Micro Motion, Inc., Boulder , Colorado. Micro Motion is a registered tra de name of Micro Motion, Inc., Boul der , Colorado. The Micro Motion and Emerson lo gos are trademarks and service ma rks of Eme rson ...

  • Emerson Process Management T-SERIES - page 3

    1 Before Y ou Begin Before Y ou Begin This manual describes ho w to install a Micro Motion ® T -Series sensor . The followi ng informat ion is provided in this manual: Customer service For technical assistance, phone the Micr o Motion Customer Service depa rtment: • In the U.S.A., phone 800-522-MASS (800-522-6277) (toll free) • In Canada and L ...

  • Emerson Process Management T-SERIES - page 4

    2 Before Y ou Begin Definitions The term MVD ™ transmitt er refers to the follo wing transmitter models: • Models 1500, 170 0, 2500, an d 2700 • Models 3500 and 3700 with sensor interf ace code 5 or 6 European installations This Micro Motion prod uct complies w ith all applicable European direct i ves when properly installed in accordance wit ...

  • Emerson Process Management T-SERIES - page 5

    3 Before Y ou Begin Figure 1 T -Series sensor with core processor Figure 2 T -Series sensor with integrally mounted Model 1700/2700 transmitter Approv al tag Flow direction arrow Core processor housing Calibration tag Process connection Approv al tag Flow direction arrow Core processor housing Calibration tag Process connection Model 1700/2700 tran ...

  • Emerson Process Management T-SERIES - page 6

    4 Determining a Location Figure 3 T -Series sensor with junction box Step 1 Determining a Location Choose a location for the sensor based on the requirements described in th is section. The foll o wing general guidelines can help you select an appropriate locatio n for the sensor . Full flow tubes For op timal performance, the sensor t ubes should ...

  • Emerson Process Management T-SERIES - page 7

    5 Determining a Locati on Figure 4 T -Series sensor ambient an d process temperat ure limits Hazardous area approv als may impose additional limits on ambient and process temperature. For the A TEX “T” rating, refer to the A TEX documentation shipped w ith the sensor or av ailable on the Micro Motion web site at www .micromotion.com . UL and CS ...

  • Emerson Process Management T-SERIES - page 8

    6 Determining a Location Maximum wiring distances If the transmitter is mounted remote ly from the sensor , the maximum di stance between the sensor and transmitter depends on cable type. See T able 2. Pipe run Micro Motion sensors d o not require a straight run of pipe upstream or downstream. Va lv e s After the sensor and transmitter ha ve been i ...

  • Emerson Process Management T-SERIES - page 9

    7 Orienting the Sensor Step 2 Orienting the Sensor The sensor will function proper ly in an y orientation if the sensor tube s remain f illed with process fluid. Micro Motion recommends orienting the sensor in a v ertical pipeline, with process flu id flo w ing upwa rd through the sensor (see Figure 5). Figure 5 Recommended se nsor orientation Self ...

  • Emerson Process Management T-SERIES - page 10

    8 Orientin g the Se nsor Figure 6 Eccentric redu cer • The gap between the ele ctronics housing and se nsor body shou ld be inspected period ically . Man ually clean this gap when necessary . • The ASME Bioprocess Equipment Standard (ASME BPE-19 97), section SC-3.12.1, includes recommendations re garding ho w to install process piping to ensure ...

  • Emerson Process Management T-SERIES - page 11

    9 Mounti ng the Sensor Step 3 Mounting the Sensor Use your commo n practices to minimize to rque and bendi ng load on process connections. T o reduce the risk of condensation or excessiv e moisture , the conduit opening should not point upward (if possible). The conduit opening of the junction box or core processo r can be rotated freely to facilit ...

  • Emerson Process Management T-SERIES - page 12

    10 Wiring Step 4 Wiring Hazardous area installations If you are installing the sensor in a hazardous location, verify that the hazardous classif ication information printed on the sensor tag matches the en vironm ent in which the sensor will be installed. Failure to comply with the requirements f or intrin sic safety in a hazardous area could resul ...

  • Emerson Process Management T-SERIES - page 13

    11 Wiring Figure 8 Micro Motion cable g land and he at shrink 4. For conn ection at the core processor housing, pr epare shielded cable as follo ws (for armored cable, omit steps d, e, f, and g): a. Strip 4 1/2 inches ( 114 mm) of c able jack et. b . Remove the clear wrap that is inside the cable jacket, and remo ve the filler material between the ...

  • Emerson Process Management T-SERIES - page 14

    12 Wiring Figure 10 Applying the heat shrink g. Position gland clamping inse rt so the interior end is flush with the heat shrink. h. Fold the cl oth shield or braid and drain wires o ve r the clamping insert and approximately 1/8 inch (3 mm) past the O-ring. Figure 11 Folding the cloth shield i. Install the gland body into the co re processor ho u ...

  • Emerson Process Management T-SERIES - page 15

    13 Wiring Figure 13 Connecting the wires at th e core processor 7. Reinstall and tighten the co re processor housing co ver . 8. Additional wiring instructions for the tran smitter can be found in the transmitter manual. Note: Never gr ound the 4-wire cable shield an d shield dr ain wir e(s) at the transmitter . Junction box to a 9-wire remote tr a ...

  • Emerson Process Management T-SERIES - page 16

    14 Grounding Step 5 Grounding The sensor can be grounde d via the piping if the joints in the pipeline are ground- bonded. If the sen sor is not grounded via the piping, connect a groun d wire to the in ternal or exter nal grounding scre w , which is located on the core processo r housing or junction box. Improper gr ounding can cause measurement e ...

  • Emerson Process Management T-SERIES - page 17

    15 Purge Fittings 4. Connect the supp ly of nitrogen or ar gon g as to the inlet pur ge connection or open inlet pur ge line. Leav e the outlet connec tion open. • Exercise caution to a voi d introducing dirt, moisture, rust, or other contaminants into the sensor case. • If the purge g as is heav ier than air (such as argo n), locate the inlet ...

  • Emerson Process Management T-SERIES - page 18

    16 Return Policy Return Policy Micro Motion procedu res must be followed when retu rning equipment. These procedures ensure le gal compliance with go vernment transpor tation agencies and help pro vide a safe working en vironment for Micro Motion employees. F ailure to follo w Micro Motion proced ures will result in your equipment being refused del ...

  • Emerson Process Management T-SERIES - page 19

    ...

  • Emerson Process Management T-SERIES - page 20

    Micro Motion Inc. USA Worldwide Headqu arters 7070 Winchester Circle Boulder , Colorado 80301 T +1 303-527-5200 +1 800-522-6277 F +1 303-530-8459 Micro Motion Europe Emerson Process Management Neonstraat 1 6718 WX Ede The Netherlands T +31 (0) 318 495 555 F +31 (0) 318 495 556 Micro Motion Japan Emerson Process Management 1-2-5, Higashi Shinagawa S ...

メーカー Emerson Process Management カテゴリー Stud Sensor

Emerson Process Management T-SERIESのメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- T-SERIESの取扱説明書
- Emerson Process Managementの製品カード
- パンフレット
- またはEmerson Process Management T-SERIESの消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Emerson Process Management T-SERIESの取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Emerson Process Management T-SERIESの取り付け説明書、サービスマニュアル、簡易説明書、またはEmerson Process Management T-SERIESのユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Emerson Process Management T-SERIESの製品を使用するにあたって最も人気のある説明書を閲覧できます。

Emerson Process Management T-SERIESデバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがEmerson Process Management T-SERIESを使用するのを助ける技術的文書のことです。説明書は通常、全てのEmerson Process Management T-SERIESユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Emerson Process Managementの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Emerson Process Management T-SERIESの説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははEmerson Process Management T-SERIESの目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Emerson Process Management T-SERIESに関してこのドキュメントで見つけることができる全てのヒントの目次
3. Emerson Process Management T-SERIESデバイスの基本機能を使うにあたってのヒント。 Emerson Process Management T-SERIESのユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Emerson Process Management T-SERIESに関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるEmerson Process Management T-SERIESのメーカー/サービスへの連絡先に関する情報。

Emerson Process Management T-SERIESについてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもEmerson Process Management T-SERIESの問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがEmerson Process Management T-SERIESで同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)