Emerson 398 - page 1
Models 396 and 398 pH/ORP Sensors Instruction Sheet PN 51A-396/rev .F January 201 1 SPECIFICA TIONS pH Range*: AccuGlass: 0-14 pH Wetted Materials: Model 396: polypropylene, 316 stainless steel, glass, EP Model 398: T efzel; titanium; glass; and choice of Kalrez, Viton or EPDM o-rings Interconnecting Cable: 15 ft (4.6 m), for use with remote preamp ...
-
Emerson 398 - page 2
STORAGE 1. It is recommended that electrodes be stored in their original shipping containers until needed. 2. Do not store at temperatures below -5°C (23°F). 3. Electrodes should be stored with a protective cap containing KCl solution (PN 9210342). 4. For overnight storage, immerse the sensor in tap water or 4 pH buffer solution. 5. A pH glass el ...
-
Emerson 398 - page 3
3 ORP CALIBRA TION 1. After making an electrical connection between the sensor and the instrument, obtain a st andard solu- tion of saturated quinhydrone. This can also be made quite simply by adding a few crystals of quin- hydrone to either pH 4 or pH 7 buffer . Quinhydrone is only slightly soluble, therefore only a few crystals will be required. ...
-
Emerson 398 - page 4
4 FIGURE 1. Sensor Dimensions MILLIMETER INCH ...
-
Emerson 398 - page 5
5 FIGURE 2. Flow Through and Insertion Inst allation For Models 396 and 398 1-1/2 inch Pipe T ee (PN 200201 1) with 1 inch threaded connections. NOTE: The sensor must be mounted at least 10° above the horizon. ...
-
Emerson 398 - page 6
6 FIGURE 3. Submersion Inst allations For Models 396 and 398 MILLIMETER INCH ...
-
Emerson 398 - page 7
7 FIGURE 4. Models 396 and 398 St andard Wiring Option used with Models 1054A, 1054B, 2054, and Preamplifiers PN 22698-02 and 22698-03 FIGURE 5. Models 396 and 398 Option -62 for Wiring to Models 1054, 1 181, 2081, 54e, 81, 3081, 4081, 5081-P , Xmt, SCL-P/Q, and 2700 Preamplifier (PN 23054-03). Also for use with remote Junction Box PN 23555-00. Fig ...
-
Emerson 398 - page 8
8 MILLIMETER INCH SENSOR MODELS 54, 54e, 3081, 81, 4081, 5081-P pH/ORP FIGURE 7. Wiring to Junction Box (PN 23555-00) and Remote Preamplifier . ...
-
Emerson 398 - page 9
9 FIGURE 10. Wiring to Model 1055-22-32. FIGURE 8. Wiring to Model Xmt FIGURE 9. Wiring to Models 3081, 4081, and 5081-P ...
-
Emerson 398 - page 10
10 FIGURE 1 1. Wiring to Model 1055-22-32 through a Remote Junction Box FIGURE 12. Wiring to Model 1055-01-10-22-32 ...
-
Emerson 398 - page 11
11 FIGURE 13. Prep aring Coax Cable with BNC Connector for Wiring to Models 54, 54e, 81, 2081, 3081, 4081, 5081-P , and Xmt-P FIGURE 14. Prep aring Coax Cable for Wiring to Model 1 181 ...
-
Emerson 398 - page 12
For more detailed instructions, see electronic file 51-396to398 on “Instruction Manuals” CD-rom. FIGURE 17. Wiring to Model 2081 FIGURE 18. Wiring to Models 1054A/B and 2054 SENSOR FIGURE 16. Wiring to Model 1 181 FIGURE 15. Wiring to Models 54e and 81 Credit Cards for U.S. Purchases Only . The right people, the right answers, right now . ON-LI ...