ResMed Positive AirwAy Pressure Device S8 Auto 25の取扱説明書

23ページ 0.58 mb
ダウンロード

ページに移動 of 23

Summary
  • ResMed Positive AirwAy Pressure Device S8 Auto 25 - page 1

    S8™ A uto 25 Positive Airw A y Pressure device User Guide English ...

  • ResMed Positive AirwAy Pressure Device S8 Auto 25 - page 2

    English S8™ A uto 25 POSITIVE AIRW A Y PRESSU RE DEVICE User Guide English ResMed Ltd (Manufacturer) 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia ResMed Corp (US Designated Agent) 14040 Danielson Street Poway CA 92064-6857 USA ResMed (UK) Ltd (EU Authorised Representative) 96 Milton Park Abingdon Oxfordshire OX14 4RY UK ResMed Offic ...

  • ResMed Positive AirwAy Pressure Device S8 Auto 25 - page 3

    Cont ents Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Contraindications 1 A dverse Ef f ects 1 The S8 Auto 25 S ystem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Masks 2 Humidifier 2 R esScan Dat a Card 2 Using the S8 Auto 25 . . . . . . . . . . . . . . . ...

  • ResMed Positive AirwAy Pressure Device S8 Auto 25 - page 4

    1 Introduction English Introduction The S8 TM Auto 25 self adjusti ng sleep apnoea system is indicated for the treatment of obstructive sle ep apnoea (OSA) in pati ents > 30 kg . The S8 Auto 25 is in tended for home and hosp ital use. Contraindications The S8 Auto 25 is not a life support device and may stop ope rating during power failure or ce ...

  • ResMed Positive AirwAy Pressure Device S8 Auto 25 - page 5

    2 Masks The following ResMed mask systems are recommende d for use with this device : For information on using masks, see your mask manual. For th e latest available masks, see www.resmed.co m. Humidifier If you are ex periencing dryness o f the nose, throat or mouth, the H4i heated humidifier is recommended for use with the S8 Auto 25. ! WARNING ? ...

  • ResMed Positive AirwAy Pressure Device S8 Auto 25 - page 6

    3 Using the S8 Auto 25 English Using the S8 Auto 25 Setting up your S8 1 Connect the power cord to the socket at the rear of your S8 Auto 25. 2 Plug the other end of the powe r cord into the power outlet. 3 Connect one e nd of the air tu bing firmly onto the air outlet. 4 Connect the assembled mask system to the free end of air tubing. For informat ...

  • ResMed Positive AirwAy Pressure Device S8 Auto 25 - page 7

    4 Attaching a H4i Humidifier The H4i humidifier attaches to th e front of the S8 Auto 25 device to pr ovide he ated humidifica tion. This device automatica lly detects the prese nce of the H4i and no othe r accessories are requi red for its use. For more information on usi ng your H4i, please refer to the H4i user guide. ! WARNING • Make sure tha ...

  • ResMed Positive AirwAy Pressure Device S8 Auto 25 - page 8

    5 Using the S8 Auto 25 English Using the Menus The S8 Auto 25 provides a set of functi ons which are arranged in menus and submenus. Via th e LCD screen, the menus and submenus allow you to view and change the settings for a particular function. To navi gate and make selections: 1 Press or to scroll through items within a lev el. 2 Press to enter a ...

  • ResMed Positive AirwAy Pressure Device S8 Auto 25 - page 9

    6 How to Select the Mask Type Scroll to MASK and sel ect . Press or until you see the setting you require. The following tabl e shows the setting that should be selected for each mask: How to change the EPR level (CPAP mode) Expiratory Pressure Reli ef (EPR) can be used to help ease any discomfort you may experience when breathing out during treatm ...

  • ResMed Positive AirwAy Pressure Device S8 Auto 25 - page 10

    7 Using the S8 Auto 25 English Starting Treatment 1 Make sure the power is on. The product name is displaye d briefly on the LCD screen, then the standby (Ramp) screen appears. The key and LCD backlights also turn on. 2 Fit your mask as described in the ma sk user instructions. ! WARNING A mask should not be use d unless your S8 Auto 25 is turne d ...

  • ResMed Positive AirwAy Pressure Device S8 Auto 25 - page 11

    8 The LCD display s a mask-fit star rating from zero to five stars. Three to five stars indicates a good fit or better. Ze ro to two stars indicates that the mask needs to be adjusted. Reminders on the S8 LCD Your clinic ian may have set your S8 Auto 25 to remind you ab out important eve nts, such as when to replace your mask, when to insert your D ...

  • ResMed Positive AirwAy Pressure Device S8 Auto 25 - page 12

    9 Using the S8 Auto 25 English Using the Data Card If your clinician needs to review your treatment, they will ask you to use the Data Card to copy data from your S8 device and return the card to them. Copying Data onto a Data Card 1 Switch on your S8 and wait unti l you see the standby (Ramp) screen. 2 Hold the Data Card with the a rrow facing up ...

  • ResMed Positive AirwAy Pressure Device S8 Auto 25 - page 13

    10 Traveling with the S8 Auto 25 International Us e Your S8 flow generator has an internal powe r adapter that enables it to operate in other countries. It will op erate on power supplies of 100–240V and 50–60Hz. No special adju stment is necessary, but you will require an approved p ower cord for that country. Use on an Aircraft Please consult ...

  • ResMed Positive AirwAy Pressure Device S8 Auto 25 - page 14

    11 Cleaning and Maintenance English Cleaning and Maintenance You should regularly carry out the cleani ng and maintenance described in this section. Refer to your mask and humidifier manual s for detailed in structions regarding the care of those devices. Daily Disconnect the air tubing and hang it in a clean , dry place until next use. Weekly 1 Re ...

  • ResMed Positive AirwAy Pressure Device S8 Auto 25 - page 15

    12 ! WARNING • Do not wash the air filter . Th e air filter is not wa shable or reusable. • The air filter co ver protects the devi ce in th e event of accidental liqu id spillage onto the de vice. Ensure the air filter and air fi lter cover are fit ted at all times. Servicing This product should be ins pected by an authorised ResMed service ce ...

  • ResMed Positive AirwAy Pressure Device S8 Auto 25 - page 16

    13 T roubleshooting English Troubleshooting If there is a problem, try th e following suggestio ns. If the problem cannot be sol ved, contact your equi pment supplier or Re sMe d. Do not attempt to open the device. No display Insufficient air delivered fr om the S8 device Device does not st art when you b rea the int o the mask Device does not s to ...

  • ResMed Positive AirwAy Pressure Device S8 Auto 25 - page 17

    14 Displays er ror message: Exxxx Call Service (wher e xxxx defines an er ror) Displays er ror message: HIGH LEAK!!! Adjust Mask The follo wing message is display ed on the LCD aft er you try to updat e settings or copy data to the D ata Car d: Card Error Remove Card The follo wing message is display ed on the LCD after y ou try to update the setti ...

  • ResMed Positive AirwAy Pressure Device S8 Auto 25 - page 18

    15 T echnical Specifications English Technical Specifications Notes: • The m anufact urer reser ves t he right to c hange these s pecificat ions without notice. • Pressure may be displayed in cm H 2 0 or hP a. Operating pressure range 4 to 25 cm H 2 O Maximum single fault steady state pressure 40 cm H 2 O Pressure measurement tolerance ±0.5 cm ...

  • ResMed Positive AirwAy Pressure Device S8 Auto 25 - page 19

    16 Symbols Which Appear On The Device Caution; Follow instruction s for use; Drip proof; Type CF equipment; Dangerous vol t age; Class II equipment; St art/Stop and mask-fit ; Manufac turer; Environmental infor mation WEEE 2002/96 /EC is a Europ ean Direct ive that re quires th e proper dis posal of elec trical an d electroni c equipment . This dev ...

  • ResMed Positive AirwAy Pressure Device S8 Auto 25 - page 20

    17 General W arnings and Cautions English Note: At a fixed rate of supplement al oxy gen flow , the inhaled oxygen concentration varies, depending on the pressu re set tings, patient breathing pat tern, mask, point of application and leak rate . • Do not use the S8 Auto 25 if there are obvious e xternal def ects or unexplained c han ges in perfor ...

  • ResMed Positive AirwAy Pressure Device S8 Auto 25 - page 21

    18 Guidance and Manuf actur er ’ s De claration—Electromagnetic Emis sions and Immunit y Guidance and manufactur er’ s declaration—electr omagnetic emissions The S8 A uto 25 d evice is i ntended fo r use in the elec tromagnetic en vironment spe cified belo w . T he customer or the user of the S8 device should assu re that the device is used ...

  • ResMed Positive AirwAy Pressure Device S8 Auto 25 - page 22

    19 General W arnings and Cautions English Guidance and manufact urer’ s declar ation—electromagnetic immunity The S8 de vice is i ntended for use in the electr omagnetic en vironment sp ecified belo w . T he customer or the us er of the S8 dev ice should assure that the device is us ed in such an environment. Immun ity test IE C 6 0 6 01- 1- 2 ...

  • ResMed Positive AirwAy Pressure Device S8 Auto 25 - page 23

    20 Limited Warranty ResMed warrants that your ResMed product shal l be free from defects in material and workmanship for the period specified below from the date o f purchase by the initial consum er. This w a rranty is not transf erable. Note: Some mod els are not av ailable in all region s. If the product fails under conditions of norm al use, Re ...

メーカー ResMed カテゴリー Respiratory Product

ResMed Positive AirwAy Pressure Device S8 Auto 25のメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- Positive AirwAy Pressure Device S8 Auto 25の取扱説明書
- ResMedの製品カード
- パンフレット
- またはResMed Positive AirwAy Pressure Device S8 Auto 25の消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、ResMed Positive AirwAy Pressure Device S8 Auto 25の取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、ResMed Positive AirwAy Pressure Device S8 Auto 25の取り付け説明書、サービスマニュアル、簡易説明書、またはResMed Positive AirwAy Pressure Device S8 Auto 25のユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、ResMed Positive AirwAy Pressure Device S8 Auto 25の製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

ResMed Positive AirwAy Pressure Device S8 Auto 25デバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがResMed Positive AirwAy Pressure Device S8 Auto 25を使用するのを助ける技術的文書のことです。説明書は通常、全てのResMed Positive AirwAy Pressure Device S8 Auto 25ユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

ResMedの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. ResMed Positive AirwAy Pressure Device S8 Auto 25の説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははResMed Positive AirwAy Pressure Device S8 Auto 25の目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。ResMed Positive AirwAy Pressure Device S8 Auto 25に関してこのドキュメントで見つけることができる全てのヒントの目次
3. ResMed Positive AirwAy Pressure Device S8 Auto 25デバイスの基本機能を使うにあたってのヒント。 ResMed Positive AirwAy Pressure Device S8 Auto 25のユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。ResMed Positive AirwAy Pressure Device S8 Auto 25に関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるResMed Positive AirwAy Pressure Device S8 Auto 25のメーカー/サービスへの連絡先に関する情報。

ResMed Positive AirwAy Pressure Device S8 Auto 25についてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもResMed Positive AirwAy Pressure Device S8 Auto 25の問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがResMed Positive AirwAy Pressure Device S8 Auto 25で同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)