    No. EJL-114B Introduction The electromagnetic flowmeter uses Faraday’s Law of electromagnetic induction to measure the process flow. The device consists of two units: a detector, through which the fluid to be measured flows and in which low-level signals proportional to flow rates are obtained; and a converter, which supplies excitation current t ...

  • Toshiba LF410/LF602 - page 2

    LF410/LF600 LF410/LF602 2 Fluid conductivity: 5µS/cm minimum Fluid temperature: –10 to +180 ° C : Ceramic type (14 to 356 ° F) Note : 120 ° C (248 ° F) above is separate type –10 to +120 ° C : tefolon PFA (14 to 248 ° F) Ambient temperature: Without cFMus Approval: - 20 to + 60 °C ( - 4 to 140 °F) With cFMus Approval: -10 to +45° C (1 ...

  • Toshiba LF410/LF602 - page 3

    LF410/LF600 LF410/LF602 3 totalizer. Fixed-value outputs — Outputs fixed-values for current and pulse outputs. Zero adjustment — Executes zero adjust ment (on-stream at zero flow rate). Output signals Current output: 4–20mAdc (load resistance 0 to 750 Ω ) Note: The current output cannot be used with the PROFIBUS-PA ccommunication. Digital ou ...

  • Toshiba LF410/LF602 - page 4

    LF410/LF600 LF410/LF602 4 Cable connection ports: Cable glands — LF600 and LF602 without cFMus Approval: Provided as standard OD of cable φ 11 ~ 13mm Material Nylon 66 G (PF) 1/2 male screws. Note: When PROFIBUS option is specified , cable grands size is φ 6 ~ 8mm for signal cable, φ 11 ~ 13mm for power cable. LF600 and LF602 with cFMus Approv ...

  • Toshiba LF410/LF602 - page 5

    LF410/LF600 LF410/LF602 5 Installation  Dimensions Unit : mm L1 Φ D1 (L2) 64 143 (35) 225 131 BS16, DIN PN16 and JIS 10K dimensions: Meter size (mm) L1 (mm) L2 (mm) D1 (mm) Weight (kg) 15 70 252 49 approx. 5 Note: 1 inch = 25.4m m ANSI class 150 and class 300 dimensions: Meter size (inch) L1 (inch) L2 (inch) D1 (inch) Weight (lb) 1/2 2.76 9.92 ...

  • Toshiba LF410/LF602 - page 6

    LF410/LF600 LF410/LF602 6 Unit : mm (L2) L1 225 133 (35) 131 Φ D1 Note1: Eye bolts are provided at the top fo r flow mete rs sized 200mm (8"). BS16, DIN PN16 and JIS 10K dimensions Meter size (mm) L1 (mm) L2 (mm) D1 (mm) Weight (kg) 40 100 264 85 approx. 6 50 110 280 102 approx. 7 80 110* 306 127 approx. 8 100 120* 338 159 approx. 10 150 230 ...

  • Toshiba LF410/LF602 - page 7

    LF410/LF600 LF410/LF602 7 Unit : mm 88 40 (126) (35) 91 R(PT) ½ Male S crew (T wo places) (L2 ) Φ 66 87 L1 Note1: Cable glands are not provide d for LF414 of FM and CSA a pproved t ype. Refer to the pa rt Cable c onnection port at det ector. Note2: 1 inch = 25.4m m BS16, DIN PN16 and JIS 10K dimensions: Meter size (mm) L1 (mm) L2 (mm) D1 (mm) Wei ...

  • Toshiba LF410/LF602 - page 8

    LF410/LF600 LF410/LF602 8 矢視A 信号ケ ー ブ ル グラ ン ド 電源ケ ー ブ ルグ ラ ン ド 封止栓 励 磁 ケー ブル グラ ンド 電流 出 力ケー ブルグラ ン ド 接地端子 (M4 ) 取付金具 4- φ 11 取付金具 82 (36) 156 13 15 1 178 200 151 74± 0.3 25 LCD 表示 器 赤外線ス イ ッ チ Exci tation ca ...

  • Toshiba LF410/LF602 - page 9

    LF410/LF600 LF410/LF602 9 T erm inal bo ard Conne cted de tecto r Excitation cabl e IV wi r e 5.5 mm 2 or more Pow e r sw itch (Ext er nal do uble -pol e po w e r sw itch ) Thi ck walled st eel con du it Signal cabl e Pow e r cable (CVV) Digital input cable (CVV -S) Curre nt output ( 4~20mA dc) or PROFI BUS Gr ounde d w ith 100 Ω or less gro und ...

  • Toshiba LF410/LF602 - page 10

    LF410/LF600 LF410/LF602 10  Wiring Precautions (1) Explosion proof type flowmeters are not provided cable glands. Refer to the part Cable connection port at detector and converter. (2) Connect the grounding wire (IV wire 5.5mm² or more) to a good earth ground (100 Ω or less ground resistance). Make the wire as short as possible. Do not use a c ...

  • Toshiba LF410/LF602 - page 11

    LF410/LF600 LF410/LF602 11  Piping Precautions (1) Design piping so that t he flowmeter detector pipe is always filled with the fluid being measured, whether the fluid is flowing or not. (2)The detector has no adjustable piping mechanism. Install an adjustable short pipe where needed. L:Required straight pipe length (3) The required straight pip ...

  • Toshiba LF410/LF602 - page 12

    LF410/LF600 LF410/LF602 12  Meter Size To select the meter size: See Figure 13 and find meter sizes within the velocity of 0.1 to 10m/s for a specified full-scale (measuring range high limit) flow. Select one that has its full-scale velocity between 1 and 3m/s. Note: Make sure the full-scale flow rate used for the final planning stage stays with ...

  • Toshiba LF410/LF602 - page 13

    LF410/LF600 LF410/LF602 13  About establishment environment Do not store or install the flowmeter in: • Places where there is direct sunlight. • Places where excessive vibration or mechanical shock occurs. • Places where high temperature or high humidity conditions obtain. • Places where corrosive atmo spheres obtain. • Places submerge ...

  • Toshiba LF410/LF602 - page 14

    LF410/LF600 LF410/LF602 14 Table 3. Specification Code (Waf er type detector LF410 Series) Detector ca tegory Model Specific ation Code Normal type Ex. type 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Description Gr.-A Gr.-B Gr .-C Gr.-D L F 4 1 0 Normal specification type √ √ L F 4 1 4 Hazardous location certification type (Note 1) √ √ M e t e r s i ...

  • Toshiba LF410/LF602 - page 15

    LF410/LF600 LF410/LF602 15 Table 4. Specification Code for converter s Model Specific ation Code 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Contents L F 6 0 Elec tromagnetic flowmeter converter LF600 type LF602 type 0 Integral (Combined) type  – 2 Separate (Remote) type –  A F Purpose Standard cFMus class I, Division 2 approved     A B ...

  • Toshiba LF410/LF602 - page 16

    LF410/LF600 LF410/LF602 No. EJL-114B Specifications are subject to change without notice. Printed in Japan 2006-10 (TDOC) © TOSHIBA Corporation 2006 All Rights Reserved. Misuse of this product can result in da mages to property or human injury. Read related manuals carefully before using this product. ISO9001 and ISO1400 1 are certified. ...

メーカー Toshiba カテゴリー Oxygen Equipment

Toshiba LF410/LF602のメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- LF410/LF602の取扱説明書
- Toshibaの製品カード
- パンフレット
- またはToshiba LF410/LF602の消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Toshiba LF410/LF602の取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Toshiba LF410/LF602の取り付け説明書、サービスマニュアル、簡易説明書、またはToshiba LF410/LF602のユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Toshiba LF410/LF602の製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Toshiba LF410/LF602デバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがToshiba LF410/LF602を使用するのを助ける技術的文書のことです。説明書は通常、全てのToshiba LF410/LF602ユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Toshibaの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Toshiba LF410/LF602の説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははToshiba LF410/LF602の目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Toshiba LF410/LF602に関してこのドキュメントで見つけることができる全てのヒントの目次
3. Toshiba LF410/LF602デバイスの基本機能を使うにあたってのヒント。 Toshiba LF410/LF602のユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Toshiba LF410/LF602に関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるToshiba LF410/LF602のメーカー/サービスへの連絡先に関する情報。

Toshiba LF410/LF602についてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもToshiba LF410/LF602の問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがToshiba LF410/LF602で同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)