  • Agilent Technologies 8753ES - page 1

    This document describes the performance and features of the following products: • Agilent 8753ES S-paramet er vector netw ork analyzer • Agilent 8753ES Option 011 vector network analyzer without test set • Agilent 8753ET T ransmission/ref lection vector netw ork analyzer F or more information about these analyzers, please see the follo wing d ...

  • Agilent Technologies 8753ES - page 2

    3 4 10 21 21 22 24 T able of Contents Definitions and T est Conditions System Performance Summaries Specifications and Characteristics Measurement Throughput Summary Options T est Set Specifications for Option 01 1 Accessories ...

  • Agilent Technologies 8753ES - page 3

    3 Specif ications describe t he instrument’s war ranted performance af ter a half-hour war m-up and ov er the temperature range of 25° ± 5° C, unless other- wise stat ed. Specifications for frequencies abo ve 3 GHz do not apply to instruments wit h Option 075 (75-ohm impedance). Supplemental charact eristics are typical but non- warranted per ...

  • Agilent Technologies 8753ES - page 4

    4 Agilent 8753ES (50-ohm systems) 7-mm test ports Configuration Network analyzer 8753ES Standard and Option 006 Calibration kit 85031B T est-port cables 11857D System dynamic range 30 kHz to 50 kHz 70 dB 1 50 kHz to 300 kHz 90 dB 1 300 kHz to 16 MHz 100 dB 16 MHz to 3 GHz 110 dB Option 014 108 dB 3 GHz to 6 GHz 105 dB Option 014 103 dB Measurement ...

  • Agilent Technologies 8753ES - page 5

    5 Agilent 8753ES (50-ohm systems) T ype-N test ports Configuration Network analyzer 8753ES Standard and Option 006 Calibration kit 85032B T est-port cables 11857D System dynamic range 30 kHz to 50 kHz 70 dB 1 50 kHz to 300 kHz 90 dB 1 300 kHz to 16 MHz 100 dB 16 MHz to 3 GHz 110 dB Option 014 108 dB 3 GHz to 6 GHz 105 dB Option 014 103 dB Measureme ...

  • Agilent Technologies 8753ES - page 6

    6 Agilent 8753ES (50-ohm systems) 3.5-mm test ports Configuration Network analyzer 8753ES Standard and Option 006 Calibration kit 85033D T est-port cables 11857D System dynamic range 30 kHz to 50 kHz 70 dB 1 50 kHz to 300 kHz 90 dB 1 300 kHz to 16 MHz 100 dB 16 MHz to 3 GHz 110 dB Option 014 108 dB 3 GHz to 6 GHz 105 dB Option 014 103 dB Measuremen ...

  • Agilent Technologies 8753ES - page 7

    Agilent 8753ES (75-Ohm Systems) T ype-N test ports Configuration Network analyzer 8753ES Option 075 Calibration kit 85036B T est-port cables 11857B System Dynamic Range 30 kHz to 50 kHz 68 dB 1 50 kHz to 300 kHz 90 dB 1 300 kHz to 16 MHz 96 dB 16 MHz to 3 GHz 106 dB Measurement Port Characteristics Frequency Range Corrected 30 kHz–300 kHz 2 300 k ...

  • Agilent Technologies 8753ES - page 8

    8 Agilent 8753ES (75-ohm systems) T ype-F test ports Configuration Network analyzer 8753E Option 075 Calibration kit 85039B T est-port cables 11857B System dynamic range 30 kHz to 50 kHz 68 dB 1 50 kHz to 300 kHz 90 dB 1 300 kHz to 16 MHz 96 dB 16 MHz to 3 GHz 106 dB Measurement Port Characteristics Data is shown for a T ype-F female reflection por ...

  • Agilent Technologies 8753ES - page 9

    9 1. One-port or enhanced-response calibration. 2. Enhanced-response or response-only calibration. 3. 2 dB less with Option 004. 4. Option 004 may degrade uncorrected source match as much as 2 dB. T ransmission uncertainty (enhanced-response calibration) T ransmission uncertainty (response calibration) Reflection uncertainty (one-port calibration) ...

  • Agilent Technologies 8753ES - page 10

    10 T est-Port Output Characteristics 7 Frequency characteristics Range 30 kHz to 3 GHz Option 006 30 kHz to 6 GHz Resolution 1 Hz Stability 9 ±7.5 ppm (0° to 55° C) ±3 ppm/year Option 1D5 ±0.05 ppm (0° to 55° C), ±0.5 ppm/year Accuracy ±10 ppm at 25° ±5° C Power range 2 -85 to +10 dBm Option 075/014 -85 to +8 dBm Power sweep range 25 dB ...

  • Agilent Technologies 8753ES - page 11

    11 External source mode 1 (CW time sweep only) Frequency range 300 kHz to 6 GHz R-channel input requirements (typical) Power level 0 to -25 dBm Spectral purity Maximum <-30 dBc spurious input Residual FM <20 kHz T ypical settling time 500 ms (automatic) 50 ms (manual) Frequency readout accuracy 0.1% (automatic) Input frequency accuracy requir ...

  • Agilent Technologies 8753ES - page 12

    12 T est Port Output Characteristics Frequency characteristics Range 300 kHz to 3 GHz Option 006 30 kHz to 6 GHz Resolution 1 Hz Stability typically ±7.5 ppm 0° to 55° C typically ±3 ppm/year Option 1D5 typically ±0.05 ppm 0° to 55° C typically ±0.5 ppm/year Accuracy ±10 ppm at 25° ± 5° C Power range -5 to +20 dBm Option 006 -5 to +18 d ...

  • Agilent Technologies 8753ES - page 13

    13 Frequency offset mode 3 Frequency range 300 kHz to 3 GHz Option 006 300 kHz to 6 GHz R-channel input requirements Power level 300 kHz to 3 GHz 0 to -35 dBm 3 GHz to 6 GHz 0 to -30 dBm LO spectral purity Maximum spurious input <-25 dBc Residual FM <20 kHz LO frequency accuracy ±1 MHz at nominal frequency External source mode 4 (CW time swe ...

  • Agilent Technologies 8753ES - page 14

    14 T est-Port Output Characteristics Frequency characteristics Range 300 kHz to 3 GHz Option 006 300 kHz to 6 GHz Resolution 1 Hz Stability 4 ±7.5 ppm (0° to 55° C) ±3 ppm/year Option 1D5 ±0.05 ppm (0° to 55° C) ±0.5 ppm/year Accuracy ±10 ppm at 25° ±5° C Power range -20 to +5 dBm -85 to +10 dBm (with Option 004) Resolution 0.01 dB Leve ...

  • Agilent Technologies 8753ES - page 15

    15 External source mode 1 (CW time sweep only) Frequency range 300 kHz to 6 GHz R-channel input requirements 2 Power level 0 to -25 dBm Spectral purity Maximum spurious input <-30 dBc Residual FM <20 kHz T ypical settling time 500 ms (automatic) 50 ms (manual) Frequency readout accuracy 2 0.1% (automatic) Input frequency accuracy requirement ...

  • Agilent Technologies 8753ES - page 16

    Measurement Number of display channels F our display channels a vailable. Number of measurement channels T w o primary and two auxiliary measurement channels available. Measurement parameters 8753ET: Ref lection, transmission, A, B, R, A/R, B/R, A/B. Conversion to im pedance or admitt ance. 8753ES: S11, S21, S12, S22, A, B, R, A/R, B/R, A/B. Conver ...

  • Agilent Technologies 8753ES - page 17

    17 Sweep trigger Set to either continuous, sing le, g roup sw eep, or external trigger . Set external trigger t o take a com- plete sweep or to measure individual points in a frequency , pow er or list sw eep. Power Set source po w er from -20 to +5 dBm f or the 8753ET 1 or from -85 to +10 dBm f or the 8753ES 2 . P o w er slope can be set in dBm/GH ...

  • Agilent Technologies 8753ES - page 18

    18 T race Functions Display data Displa y current measurement data, memor y dat a, or current measurement wit h measurement and memor y dat a simultaneously. T race math V ect or division or subtraction of cur rent linear measurement values and memor y dat a. Display annotations Star t/st op, center/span, or CW frequency , source lev el, scale/div ...

  • Agilent Technologies 8753ES - page 19

    19 Frequency offset on/off Sets the RF source to be swept at a fixed offset frequency abo ve the receiver as required in a sw ept RF/IF , fixed LO, mix er t est. Service menu Select the desired service test, service diagnostic, service or verification mode. T est Sequences Description Creat e, edit, sav e, or recall a series of front-panel k ey- s ...

  • Agilent Technologies 8753ES - page 20

    20 Front Panel Connectors 8753ES test ports (without Option 011) Connector type 7 mm, precision Impedance 50 ohms (nominal) 8753ES Option 011 test ports Connector type T ype-N female Impedance 50 ohms (nominal) 8753ES Option 075 test ports Connector type T ype-N female Impedance 75 ohms (nominal) 8753ET test ports Connector type T ype-N female Impe ...

  • Agilent Technologies 8753ES - page 21

    21 General Conditions RFI and EMI susceptibility: defined by VDE 0730, CISPR Publication 11, and FCC Class B Standards. ESD (electrostatic discharge): must be eliminated by use of static-safe w ork procedures and an anti- static bench mat. The f lexible r ubber ke ypad pro- tects ke y cont acts from dust, but the environment should be as dust-free ...

  • Agilent Technologies 8753ES - page 22

    22 Agilent 85046A/B S-Parameter T est Sets The 85046A/B S-parameter test sets provide the capability to measure ref lection and transmission characteristics (including S-parameters) of two- port devices in either direction wit h a sing le connection. The test sets are controlled from the 8753ES Option 011 and include a programmable step attenuat or ...

  • Agilent Technologies 8753ES - page 23

    23 Agilent 85047A S-Parameter T est Set The 85047A S-parameter test set provides the capability to simult aneously measure the ref lec- tion and transmission characteristics of two-port devices in either direction with a sing le connec- tion. This test set includes a frequency doubler that can be swit ched in by an 8753B/C Option 006 to measure 3 M ...

  • Agilent Technologies 8753ES - page 24

    24 Calibration Kits V ect or accuracy enhancement procedures require that t he systematic errors of the measurement sys- t em be characterized b y measuring kno wn devices (standards) on the syst em ov er the frequency range of interest. Return loss specif ications or typical values are pro vided where a vailable for the ter minations and adapters. ...

  • Agilent Technologies 8753ES - page 25

    25 Agilent 85038 Family of Calibration Kits The 85038A calibration kit contains open circuits, short circuits, and ter minations with bot h male and female 7-16 connectors. More economical v er- sions of this kit are available: the 85038M contains only male de vices, and the 85038F provides onl y female de vices (see f ollo wing t able). Each kit c ...

  • Agilent Technologies 8753ES - page 26

    26 V erification Kits Measuring kno wn devices other than t he st andards used in calibration is an easy way to v erify the proper operation of an 8753ES measurement sys- t em. Agilent of fers v erification kits which include devices, with data, for verifying the error -correct ed measurements of an 8753ES and 50-ohm test sets. 85029B 7-mm verifica ...

  • Agilent Technologies 8753ES - page 27

    27 Connectors 50-ohm type-N (f) to 75-ohm type-N (m) standard Option 004, 50-ohm type-N (m) to 75-ohm type-N (f) Dimensions 14-mm D x 70-mm L (0.56 in x 2.75 in) W eight Net 0.1 kg (0.316 lb) 50-Ohm Accessory Kits The 11853A 50-ohm type-N and the 11854A 50-ohm BNC accessory kits provide the RF com ponents generally required when using either the 85 ...

  • Agilent Technologies 8753ES - page 28

    Agilent 8347A RF Amplifier The 8347A RF amplifier delivers increased po wer acr oss a 300 kHz to 3 GHz frequency range. The 8347A pro vides lev eled output po w er without using an exter nal coupler and detect or , since these par ts are built-in. The exter nal ALC can be directly connected to the ext ernal AM input on the 8753ET/ES. This capabilit ...

Manufacturer Agilent Technologies Category Network Router

Documents that we receive from a manufacturer of a Agilent Technologies 8753ES can be divided into several groups. They are, among others:
- Agilent Technologies technical drawings
- 8753ES manuals
- Agilent Technologies product data sheets
- information booklets
- or energy labels Agilent Technologies 8753ES
All of them are important, but the most important information from the point of view of use of the device are in the user manual Agilent Technologies 8753ES.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Agilent Technologies 8753ES, service manual, brief instructions and user manuals Agilent Technologies 8753ES. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Agilent Technologies 8753ES.

A complete manual for the device Agilent Technologies 8753ES, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Agilent Technologies 8753ES by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Agilent Technologies 8753ES.

A complete Agilent Technologies manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Agilent Technologies 8753ES - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Agilent Technologies 8753ES, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Agilent Technologies 8753ES, that we can find in the current document
3. Tips how to use the basic functions of the device Agilent Technologies 8753ES - which should help us in our first steps of using Agilent Technologies 8753ES
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Agilent Technologies 8753ES
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Agilent Technologies 8753ES in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Agilent Technologies 8753ES?

Use the form below

If you did not solve your problem by using a manual Agilent Technologies 8753ES, ask a question using the form below. If a user had a similar problem with Agilent Technologies 8753ES it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)