    USER’S GUIDE M EDICAL PRODUCT INNOV A TIONS SINCE 192 8 o ™ M ODELS CA-3000 & CA-3200 o M ODELS CM-3000 & CM-3200 92804 3/28/01 8:35 AM Page 2 ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 2

    1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Warnings and Cautions . . . . . . . . . . . . . . . . . . 3 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . 4 Controls, Connectors and Visual Indicators . . . . 5 Operating Procedure . . . . . . . . . . . . . . . . . . . . 6 Cleaning and Disinfection . . . . . . ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 3

    The Emerson CoughAssist™ Mechanical In-Exsufflator (MI-E) assists patients in clearing retained bronchopulmonary secretions by gradually applying a positive pressure to the airway , then rapidly shifting to a negative pressure. This rapid shift in pressure, via a facemask, mouthpiece or an endotracheal or tracheostomy tube, produces a high expira ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 4

    Class 1 Equipment: Equipment in which protection against electric shock does not rely on basic insulation only , but includes a grounding pin on the power cord. For ground reliability always plug the power cord into an AC grounded outlet. T ype BF Equipment: T ype B piece of equipment with an F-T ype applied part. A T ype B piece of equipment is on ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 5

    D EFINITIONS Throughout this guide the following definitions apply: • Warning/DANGER: A condition that could cause electrocution or injury to a user or operator if instructions are not followed. • CAUTIONS: A condition that could cause damage to equipment or cause inaccurate function. W ARNINGS V Patients known to have cardiac instability shoul ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 6

    V Replace fuses only with ones having the same ratings for blow characteristics, current and voltage. V Do not remove the cover; there are no serviceable parts inside the unit. Refer all service to authorized personnel. V Always check time and pressure settings before each treatment. C AUTIONS • Federal Law (USA) restricts this device to use by , ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 7

    60 cm H 2 O (44 mm Hg) 60 cm H 2 O (44 mm Hg) Conforms to UL STD 2601-1, certified to CAN/CSA - STD C22.2 No. 601.1-M90, 0413 3.3 liters/second with inhale flow set to minimum; if set to maximum inhalation, the flow is the same as the exhalation flow 10 liters/second; actual flow depends on maximum pressure and on airway resistance -70 to 0 to +70 ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 8

    CA-3000, CA-3200 F RONT P ANEL C ONTROLS 1. Manual/Auto: Changes the cycling mechanism to either manual or auto- matic mode. 2. Inhale : Sets time interval (sec) for Inhale phase of automatic cycling. Not operative in the manual mode. 3. Exhale: Sets time interval (sec) for Exhale phase of automatic cycling. Not operative in the manual mode. 4. Pau ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 9

    1. Pressure: Varies the inhalation and exhalation pressures together (also see Inhale Pressure). 2. Manual Control Lever: Use to manually cycle the unit to inhale or exhale. 3. Power: The power switch turns on or off the unit. The “ I ” symbol on the switch designates the on position. When activated a green light within the switch will illumina ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 10

    B ACK P ANEL ( ALL MODELS ) 1. Cord Wrap/Breathing Hose Holder 2. Power Cord Receptacle: Securely connects the power cord to the receptacle. 3. Replacement fuse location 4. This symbol, located next to the protective earth terminal inside the unit, signifies that the device is earthed (grounded). 5. This symbol signifies the year of manufacture of ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 11

    I NITIAL S ET - UP 1. Install the power cord right angle connector to the receptacle on the rear of the device. Run the cord inside of the lower cord wrap to act as a strain relief. 2. Position the unit on a suitable surface within easy reach of the patient, or the operator of the unit. CAUTION: Position the device so that the air intake ports on t ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 12

    P RESSURE A DJUSTMENT 1. T urn on the power switch. 2. Set the inhale flow to full or reduced. 3. Attach the patient circuit to the unit and block the end of the breathing hose. 4. Set the manual/auto switch to manual (automatic models only). 5. Push the manual control lever to the exhalation phase (to the left). Observe the pressure gauge on the d ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 13

    T IMING A DJUSTMENT (CA-3000, CA-3200 MODELS ONL Y ) If the automatic feature of the CoughAssist MI-E is to be used, adjust the times as follows: 1. Each cough cycle consists of an inhalation phase, an exhalation phase and a pause phase, after which inhalation begins again. The time for each phase is set with the three knobs on the left side of the ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 14

    Automatic Operation (CA-3000, CA-3200 models only): 1. Attach the appropriate interface to the patient. 2. T o operate the unit automatically , set the manual/auto switch to the auto position. The unit will cycle from inhale (positive) to exhale (negative) to zero pressure, and back to inhale. 3. After 4 to 5 cycles, set the manual/auto switch back ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 15

    P A TIENT C IRCUIT Institutional (hospital) Use: 1. Br eathing Hose, Patient Interface and Adapters: If the device is to be used by more than one patient, the circuit must be replaced. 2. Bacterial/V iral Filter: If the device is to be used by more than one patient the filter must be replaced to prevent cross contamination. Do not try to wash the f ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 16

    SECTION 7 A CCESSORIES CoughAssist MI-E Patient Circuit The CoughAssist MI-E Patient Circuit consists o f one 3-foot long flexible smooth bore tube, a bacterial/viral filter , an adult face- mask and an adapter . Breathing Hose 3-foot long flexible smooth bore tubing with 22mm interior diameter . NOTE: Corrugated tubing may cause a small reduction ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 17

    P REVENTIVE M AINTENANCE This device has been designed to provide virtually maintenance-free operation for extended periods of time. Sharp blows to the unit or dropping the unit is to be avoided. No routine maintenance is required. G ENERAL S UGGESTIONS 1. Keep the unit’ s exterior clean. 2. Check that the air intake ports are not blocked. 3. Kee ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 18

    T ROUBLESHOOTING G UIDE 1. Pressure Gauge Adjustment: If the pressure gauge does not go to “0” when the device is turned off it must be readjusted. Remove the adhesive cover over the pressure gauge zero adjust and turn the adjust screw to “0” using a screwdriver . If unable to make this adjustment, the unit should be returned for servicing. ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 19

    C USTOMER S ER VICE : For questions about this product please con- tact Customer Service at: J.H. Emerson Co. 22 Cottage Park Avenue Cambridge, MA 02140-1691 Phone: 800-252-1414 or 617-864-1414 Fax: 617-868-0841 Email: info@jhemerson.com Web: www .coughassist.com AUTHORIZED REPRESENT A TIVE FOR THE EUROPEAN COMMUNITY : Emergo Europe P .O. Box 149 4 ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 20

    J.H. E MERSON C O . 22 Cottage Park Avenue Cambridge, MA 02140-1691 Phone: 800-252-1414 or 617-864-1414 Fax: 617-868-0841 Email: info@jhemerson.com Web: www .coughassist.com 910-1111-2 92804 3/28/01 8:35 AM Page 1 ...

Manufacturer Emerson Category Heat Pump

Documents that we receive from a manufacturer of a Emerson CA-3000, CA-3200, CM-3000, CM-3200 can be divided into several groups. They are, among others:
- Emerson technical drawings
- CA-3000, CA-3200, CM-3000, CM-3200 manuals
- Emerson product data sheets
- information booklets
- or energy labels Emerson CA-3000, CA-3200, CM-3000, CM-3200
All of them are important, but the most important information from the point of view of use of the device are in the user manual Emerson CA-3000, CA-3200, CM-3000, CM-3200.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Emerson CA-3000, CA-3200, CM-3000, CM-3200, service manual, brief instructions and user manuals Emerson CA-3000, CA-3200, CM-3000, CM-3200. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Emerson CA-3000, CA-3200, CM-3000, CM-3200.

Similar manuals

A complete manual for the device Emerson CA-3000, CA-3200, CM-3000, CM-3200, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Emerson CA-3000, CA-3200, CM-3000, CM-3200 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Emerson CA-3000, CA-3200, CM-3000, CM-3200.

A complete Emerson manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Emerson CA-3000, CA-3200, CM-3000, CM-3200 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Emerson CA-3000, CA-3200, CM-3000, CM-3200, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Emerson CA-3000, CA-3200, CM-3000, CM-3200, that we can find in the current document
3. Tips how to use the basic functions of the device Emerson CA-3000, CA-3200, CM-3000, CM-3200 - which should help us in our first steps of using Emerson CA-3000, CA-3200, CM-3000, CM-3200
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Emerson CA-3000, CA-3200, CM-3000, CM-3200
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Emerson CA-3000, CA-3200, CM-3000, CM-3200 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Emerson CA-3000, CA-3200, CM-3000, CM-3200?

Use the form below

If you did not solve your problem by using a manual Emerson CA-3000, CA-3200, CM-3000, CM-3200, ask a question using the form below. If a user had a similar problem with Emerson CA-3000, CA-3200, CM-3000, CM-3200 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)