    USER’S GUIDE M EDICAL PRODUCT INNOV A TIONS SINCE 192 8 o ™ M ODELS CA-3000 & CA-3200 o M ODELS CM-3000 & CM-3200 92804 3/28/01 8:35 AM Page 2 ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 2

    1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Warnings and Cautions . . . . . . . . . . . . . . . . . . 3 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . 4 Controls, Connectors and Visual Indicators . . . . 5 Operating Procedure . . . . . . . . . . . . . . . . . . . . 6 Cleaning and Disinfection . . . . . . ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 3

    The Emerson CoughAssist™ Mechanical In-Exsufflator (MI-E) assists patients in clearing retained bronchopulmonary secretions by gradually applying a positive pressure to the airway , then rapidly shifting to a negative pressure. This rapid shift in pressure, via a facemask, mouthpiece or an endotracheal or tracheostomy tube, produces a high expira ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 4

    Class 1 Equipment: Equipment in which protection against electric shock does not rely on basic insulation only , but includes a grounding pin on the power cord. For ground reliability always plug the power cord into an AC grounded outlet. T ype BF Equipment: T ype B piece of equipment with an F-T ype applied part. A T ype B piece of equipment is on ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 5

    D EFINITIONS Throughout this guide the following definitions apply: • Warning/DANGER: A condition that could cause electrocution or injury to a user or operator if instructions are not followed. • CAUTIONS: A condition that could cause damage to equipment or cause inaccurate function. W ARNINGS V Patients known to have cardiac instability shoul ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 6

    V Replace fuses only with ones having the same ratings for blow characteristics, current and voltage. V Do not remove the cover; there are no serviceable parts inside the unit. Refer all service to authorized personnel. V Always check time and pressure settings before each treatment. C AUTIONS • Federal Law (USA) restricts this device to use by , ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 7

    60 cm H 2 O (44 mm Hg) 60 cm H 2 O (44 mm Hg) Conforms to UL STD 2601-1, certified to CAN/CSA - STD C22.2 No. 601.1-M90, 0413 3.3 liters/second with inhale flow set to minimum; if set to maximum inhalation, the flow is the same as the exhalation flow 10 liters/second; actual flow depends on maximum pressure and on airway resistance -70 to 0 to +70 ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 8

    CA-3000, CA-3200 F RONT P ANEL C ONTROLS 1. Manual/Auto: Changes the cycling mechanism to either manual or auto- matic mode. 2. Inhale : Sets time interval (sec) for Inhale phase of automatic cycling. Not operative in the manual mode. 3. Exhale: Sets time interval (sec) for Exhale phase of automatic cycling. Not operative in the manual mode. 4. Pau ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 9

    1. Pressure: Varies the inhalation and exhalation pressures together (also see Inhale Pressure). 2. Manual Control Lever: Use to manually cycle the unit to inhale or exhale. 3. Power: The power switch turns on or off the unit. The “ I ” symbol on the switch designates the on position. When activated a green light within the switch will illumina ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 10

    B ACK P ANEL ( ALL MODELS ) 1. Cord Wrap/Breathing Hose Holder 2. Power Cord Receptacle: Securely connects the power cord to the receptacle. 3. Replacement fuse location 4. This symbol, located next to the protective earth terminal inside the unit, signifies that the device is earthed (grounded). 5. This symbol signifies the year of manufacture of ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 11

    I NITIAL S ET - UP 1. Install the power cord right angle connector to the receptacle on the rear of the device. Run the cord inside of the lower cord wrap to act as a strain relief. 2. Position the unit on a suitable surface within easy reach of the patient, or the operator of the unit. CAUTION: Position the device so that the air intake ports on t ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 12

    P RESSURE A DJUSTMENT 1. T urn on the power switch. 2. Set the inhale flow to full or reduced. 3. Attach the patient circuit to the unit and block the end of the breathing hose. 4. Set the manual/auto switch to manual (automatic models only). 5. Push the manual control lever to the exhalation phase (to the left). Observe the pressure gauge on the d ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 13

    T IMING A DJUSTMENT (CA-3000, CA-3200 MODELS ONL Y ) If the automatic feature of the CoughAssist MI-E is to be used, adjust the times as follows: 1. Each cough cycle consists of an inhalation phase, an exhalation phase and a pause phase, after which inhalation begins again. The time for each phase is set with the three knobs on the left side of the ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 14

    Automatic Operation (CA-3000, CA-3200 models only): 1. Attach the appropriate interface to the patient. 2. T o operate the unit automatically , set the manual/auto switch to the auto position. The unit will cycle from inhale (positive) to exhale (negative) to zero pressure, and back to inhale. 3. After 4 to 5 cycles, set the manual/auto switch back ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 15

    P A TIENT C IRCUIT Institutional (hospital) Use: 1. Br eathing Hose, Patient Interface and Adapters: If the device is to be used by more than one patient, the circuit must be replaced. 2. Bacterial/V iral Filter: If the device is to be used by more than one patient the filter must be replaced to prevent cross contamination. Do not try to wash the f ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 16

    SECTION 7 A CCESSORIES CoughAssist MI-E Patient Circuit The CoughAssist MI-E Patient Circuit consists o f one 3-foot long flexible smooth bore tube, a bacterial/viral filter , an adult face- mask and an adapter . Breathing Hose 3-foot long flexible smooth bore tubing with 22mm interior diameter . NOTE: Corrugated tubing may cause a small reduction ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 17

    P REVENTIVE M AINTENANCE This device has been designed to provide virtually maintenance-free operation for extended periods of time. Sharp blows to the unit or dropping the unit is to be avoided. No routine maintenance is required. G ENERAL S UGGESTIONS 1. Keep the unit’ s exterior clean. 2. Check that the air intake ports are not blocked. 3. Kee ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 18

    T ROUBLESHOOTING G UIDE 1. Pressure Gauge Adjustment: If the pressure gauge does not go to “0” when the device is turned off it must be readjusted. Remove the adhesive cover over the pressure gauge zero adjust and turn the adjust screw to “0” using a screwdriver . If unable to make this adjustment, the unit should be returned for servicing. ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 19

    C USTOMER S ER VICE : For questions about this product please con- tact Customer Service at: J.H. Emerson Co. 22 Cottage Park Avenue Cambridge, MA 02140-1691 Phone: 800-252-1414 or 617-864-1414 Fax: 617-868-0841 Email: info@jhemerson.com Web: www .coughassist.com AUTHORIZED REPRESENT A TIVE FOR THE EUROPEAN COMMUNITY : Emergo Europe P .O. Box 149 4 ...

  • Emerson CA-3000, CA-3200, CM-3000, CM-3200 - page 20

    J.H. E MERSON C O . 22 Cottage Park Avenue Cambridge, MA 02140-1691 Phone: 800-252-1414 or 617-864-1414 Fax: 617-868-0841 Email: info@jhemerson.com Web: www .coughassist.com 910-1111-2 92804 3/28/01 8:35 AM Page 1 ...

Производитель Emerson Категория Heat Pump

Документы, которые мы получаем от производителя устройства Emerson CA-3000, CA-3200, CM-3000, CM-3200 мы можем разделить на несколько групп. Это в частности:
- технические чертежи Emerson
- инструкции обслуживания CA-3000, CA-3200, CM-3000, CM-3200
- паспорта изделия Emerson
- информационные брошюры
- энергетические этикетки Emerson CA-3000, CA-3200, CM-3000, CM-3200
Все из них важны, однако самую важную информацию с точки зрения пользователя мы найдем в инструкции обслуживания Emerson CA-3000, CA-3200, CM-3000, CM-3200.

Группа документов, определяемая как инструкции обслуживания, делится также на более подробные типы, такие как: Инструкции монтажа Emerson CA-3000, CA-3200, CM-3000, CM-3200, инструкции обслуживания, короткие инструкции или инструкции пользователя Emerson CA-3000, CA-3200, CM-3000, CM-3200. В зависимости от потребностей, Вам необходимо поискать требуемый документ. На нашем сайте Вы можете просмотреть самую популярную инструкцию использования изделия Emerson CA-3000, CA-3200, CM-3000, CM-3200.

Похожие инструкции обслуживания

Полная инструкция обслуживания устройства Emerson CA-3000, CA-3200, CM-3000, CM-3200, как должна выглядеть?
Инструкция обслуживания, определяемая также как пособие пользователя, или просто "руководство" - это технический документ, цель которого заключается в использовании Emerson CA-3000, CA-3200, CM-3000, CM-3200 пользователями. Инструкции пишет, как правило технический писатель, языком, доступным для всех пользователей Emerson CA-3000, CA-3200, CM-3000, CM-3200.

Полная инструкция обслуживания Emerson, должна заключать несколько основных элементов. Часть из них менее важная, как например: обложка / титульный лист или авторские страницы. Однако остальная часть, должна дать нам важную с точки зрения пользователя информацию.

1. Вступление и рекомендации, как пользоваться инструкцией Emerson CA-3000, CA-3200, CM-3000, CM-3200 - В начале каждой инструкции, необходимо найти указания, как пользоваться данным пособием. Здесь должна находится информация, касающаяся местонахождения содержания Emerson CA-3000, CA-3200, CM-3000, CM-3200, FAQ и самых распространенных проблем - то есть мест, которые чаще всего ищут пользователи в каждой инструкции обслуживания
2. Содержание - индекс всех советов, касающихся Emerson CA-3000, CA-3200, CM-3000, CM-3200, которое найдем в данном документе
3. Советы по использованию основных функций устройства Emerson CA-3000, CA-3200, CM-3000, CM-3200 - которые должны облегчить нам первые шаги во время использования Emerson CA-3000, CA-3200, CM-3000, CM-3200
4. Troubleshooting - систематизированный ряд действия, который поможет нам диагностировать а в дальнейшем очередность решения важнейших проблем Emerson CA-3000, CA-3200, CM-3000, CM-3200
5. FAQ - чаще всего задаваемые вопросы
6. Контактные данные Информация о том, где искать контактные данные производителя / сервисного центра Emerson CA-3000, CA-3200, CM-3000, CM-3200 в данной стране, если самостоятельно не получится решить проблему.

У вас вопрос, касающийся Emerson CA-3000, CA-3200, CM-3000, CM-3200?

Воспользуйтесь формуляром, находящимся ниже

Если с помощью найденной инструкции Вы не решили свою проблему с Emerson CA-3000, CA-3200, CM-3000, CM-3200, задайте вопрос, заполнив следующий формуляр. Если у какого то из пользователей была похожая проблема с Emerson CA-3000, CA-3200, CM-3000, CM-3200 со всей вероятностью он захочет поделиться методом ее решения.

Перепишите текст с картинки

Комментарии (0)