    Model 5081T -FF/FI Instruction Sheet PN 51A-5081T -FF/FI/rev .A November 2005 Y our purchase from Rosemount Analytical, Inc. has result- ed in one of the finest instruments available for your p ar- ticular application. These instrument s have been designed, and tested to meet many national and interna- tional standards. Experience indicates that it ...

  • Emerson 5081T-FF/FI - page 2

    PHYSICAL SPECIFICA TIONS HOUSING: Epoxy-polyester painted over low-copper alu- minum. Neoprene O-rings on cover . 160.5 mm x 175.3 mm x 161.3 mm (6.3 in. x 6.9 in. x 6.4 in.) DIAMETER: 155.4 mm (6.1 in.) ELECTRICAL CONDUIT OPENINGS: 3/4 in. FNPT POWER SUPPL Y AND LOAD: A power supply voltage of 9 Vdc to 32 Vdc at 22 mA is required; Intrinsically Sa ...

  • Emerson 5081T-FF/FI - page 3

    3 3 MODEL 5081T -FF/FI SPECIFICA TIONS TRANSMITTER DISPLA Y DURING CALIBRA TION AND PROGRAMMING (FIGURE 1) 1. Continuous display of conductivity , % concentration, or custom reading. 2. Units: µS/cm, mS/cm, ppm, %, or none. 3. Current menu appears here. 4. Submenus, prompts, and diagnostic readings appear hear . 5. Commands available in each subme ...

  • Emerson 5081T-FF/FI - page 4

    4 U1 7 :! <[9:1 -E< / ! 922<U '"*#=22 /1[[! /9P<U <Q2U 4-SG U<[9:1 2D>[=T '""" 25.0C 12.00mA CAL key PROG key DIAG key HOLD key Model 5081T -FF/FI Process Display Screen µS/cm /-[5ES-U1 /1[[! /9P<U <1P<9S! " U1 7 :! -G6 U1 7 : G5<:[-A <1U>:! /V<= G12->[= PROGRAM Proc ...

  • Emerson 5081T-FF/FI - page 5

    MODEL 5081T -FF/FI DEF AUL T SETTINGS V ARIABLE NAME MNEMONIC F ACTOR Y SETTINGS CUSTOMER SETTINGS Program Menu T emperature UH OR Auto temperature compensation UDVUQ on ___________ Manual temperature UODP 25.0°C (overridden by auto) ___________ T emperature compensation algorithm /97: ( [LPHDS or P9P1 ) LInEAr ___________ Display GLTRND Y Measure ...

  • Emerson 5081T-FF/FI - page 6

    6 MODEL 5081T -FF/FI WIRING WIRING Wire sensor as shown below in Figure 2. Keep sensor wiring separate from power wiring. For best EMI/RFI protection, use shielded output signal cable in an earth-grounded metal conduit. Refer to the sensor instruction manual for more det ails. FIGURE 2. Wiring Model 5081T -FF/FI ...

  • Emerson 5081T-FF/FI - page 7

    7 7 MODEL 5081T -FF/FI WIRING WIRING THROUGH A JUNCTION BOX The sensor can be wired to the analyzer through a remote junction box (PN 23550-00). Wire the extension cable and sensor cable point-to-point. Refer to the sensor instruction manual for more details. Factory-terminated (PN 23747-00) and unterminated (PN 9200275) connecting cable are availa ...

  • Emerson 5081T-FF/FI - page 8

    8 MODEL 5081T -FF/FI INST ALLA TION INST ALLA TION UNP ACKING AND INSPECTION Inspect the shipping container . If it is damaged, cont act the shipper immediately for instructions. Save the box. If there is no apparent damage, unp ack the container . Be sure all items shown on the packing list are present. If items are missing, notify Rosemount Analy ...

  • Emerson 5081T-FF/FI - page 9

    9 ...

  • Emerson 5081T-FF/FI - page 10

    10 ...

  • Emerson 5081T-FF/FI - page 11

    11 ...

  • Emerson 5081T-FF/FI - page 12

    1 1 2 2 ...

  • Emerson 5081T-FF/FI - page 13

    13 ...

  • Emerson 5081T-FF/FI - page 14

    1 1 4 4 FM A 5/6/04 J. FLOCK 5/6/04 J. FLOCK .140 ±. 0 0 5 1.30 ±. 0 2 .650 ±. 0 15 .125 2.180 ±. 0 0 5 2.56 ±. 0 2 .120 ±. 0 15 Ø .125 R .25 5-6-04 89 33 4 4 FINISH:SILKSCREEN BLACK EPOX Y PAINT (BAKED). 5081-T-FI-67 MODEL 5081-T-FI HARDNES S BRINEL L 190 . BE ANNEALED & PASSIVATED. MAX IMUM STEEL .015+/-.005 THICK. MATERIAL TO 1 MATERI ...

  • Emerson 5081T-FF/FI - page 15

    1 1 5 5 5/6/04 J. FLOCK 5/6/04 J. FLOCK 70°C MA X CLASS III, DI V 1 CLASS II, DIV 1, GPS E-G CLASS I, DIV 1, GPS B-D HAZARDOU S AREA 3 R T D C O M 6 R C V S H L D 7 R C V C O M 8 R C V 9 D R V S H L D 1 0 D R V C OM 2 R T D S H L D 5 R T D I N 1 2 N C 1 3 N C 4 R T D S E N S E 1 R E S E R V E D 1 5 H T / F F ( - ) 1 6 H T / F F ( + ) 1 4 N C 1 1 D ...

  • Emerson 5081T-FF/FI - page 16

    1 1 6 6 FM B La, Lt OR Lo; OR Lc/Rc (La/Ra OR Lo/Ro) AND Li/Ri (L a/Ra OR Lo /Ro) 1. INSTALLATION SHOULD BE IN ACCORDANC E W ITH ANSI /ISA RP12.06 "INSTALLATION OF INTRIN SICALLY SAFE SYSTEMS FOR H AZARD OUS (C LASSIFIED) LOC ATIONS" (EXCE PT CHAPTER 5 FOR FISCO INSTALLATIONS) AND THE NATIONAL ELECTRICAL CODE (A NSI/NFPA 70) SECTIONS 504 ...

  • Emerson 5081T-FF/FI - page 17

    1 1 7 7 THIS DOCUMENT IS CERTIFIED BY REV REV REV REV REV REV REVISIONS NOT PERMITTED W/O AGENCY APPROVAL DIMENSIONS ARE IN INC HES REMOVE BURRS & SHA RP EDGES .020 MAX MACHINED FILLE T RADII .020 MAX - ITEM PART NO DESCRIPTION QT Y SCALE SIZE REV DW G NO BILL OF M AT ERIAL MATERIAL FINIS H DRAW N CHECKED PROJECT ENGR APVD NOTES: UNLESS OTHERW ...

  • Emerson 5081T-FF/FI - page 18

    1 1 8 8 Rosemount Analytical, Uniloc Divisi on 2400 Barran ca Pkwy Irvine, CA 92 606 UNLESS OTHERWISE SPECIFIED DIMENSIO NS ARE IN INCHES REMOVE BURRS & S HARP EDGES .02 0 MAX MACHINED F ILLET RADII .0 20 MAX FINISH MATERIAL PROJECT ENGR A PVD SIZE DWG NO. REV TITLE SCALE TYPE This document contains information propriet ary to Rosemount Analyti ...

  • Emerson 5081T-FF/FI - page 19

    1 1 9 9 9241514-01/ A DIME NSIONS ARE IN I NCHES REMOVE BURRS & SHA RP EDGES .020 MAX MACHINED FILLE T RADII .020 MA X - ITEM PART NO DESCRIPTION QT Y SCALE SIZE REV DW G NO BILL OF M AT ERIAL MATERIAL FINIS H DRAW N CHECKED PROJECT ENGR APVD NOTES: UNLESS OTHERW ISE SPECIFIED REV ECO NO RELEASE DATE LTR ECO REVISIONS DESCRIPTION BY DATE CHK Th ...

  • Emerson 5081T-FF/FI - page 20

    2 2 0 0 Rosemount Analytical, Uniloc Division 2400 Ba rranca Pk wy Irvine, CA 9 2 606 UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE IN INCHES REMOVE BURRS & SHARP EDGES .020 MAX MACHINED FILLET RADII .020 MAX FINISH MATER IAL PROJECT ENGR APV D SIZE DWG NO. REV TITLE SCALE TYPE This document contains information proprietary to Rosemount Analytical, ...

  • Emerson 5081T-FF/FI - page 21

    2 2 1 1 OVERVIEW This section gives general procedures for routine maintenance of the 5081-T transmitter . The transmitter needs almost no routine maintenance. TRANSMITTER MAINTENANCE Periodically clean the transmitter window with household ammonia or glass cleaner . The detector for the infrared remote controller is located behind the window at th ...

  • Emerson 5081T-FF/FI - page 22

    2 2 2 2 Replacement Part s for Model 5081-T -FF/FI T ransmitter Location in Shipping drawing PN Description Weight 1 23992-07 PCB stack consisting of the CPU (p art 3) and analog (part 4) boards, 1 lb/0.5 kg display board is not included, CPU and analog boards are factory- calibrated as a unit and cannot be ordered separately 2 23638-01 LCD display ...

  • Emerson 5081T-FF/FI - page 23

    2 2 3 3 QUICK-ST ART 1. On the Remote, press PROG , NEXT , ENTER. 2. Use the arrow buttons to select COnduc (conductivity), nAOH (Sodium Hydroxide 0-15%), HCL (Hydrochloric Acid 0-16%), H2SO4L (Sulfuric Acid 0-30%), H2SO4H (Sulfuric Acid 95-99.99%), or CuSt (custom curve) mode. Press ENTER. If you chose CuSt , continue with step 3. If you chose COn ...

  • Emerson 5081T-FF/FI - page 24

    Credit Cards for U.S. Purchases Only . The right people, the right answers, right now . ON-LINE ORDERING NOW A V AILABLE ON OUR WEB SITE http://www .raihome.com Specifications subject to change without notice. Emerson Process Management Liquid Division 2400 Barranca Parkway Irvine, CA 92606 USA T el: (949) 757-8500 Fax: (949) 474-7250 http://www .r ...

Manufacturer Emerson Category TV Converter Box

Documents that we receive from a manufacturer of a Emerson 5081T-FF/FI can be divided into several groups. They are, among others:
- Emerson technical drawings
- 5081T-FF/FI manuals
- Emerson product data sheets
- information booklets
- or energy labels Emerson 5081T-FF/FI
All of them are important, but the most important information from the point of view of use of the device are in the user manual Emerson 5081T-FF/FI.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Emerson 5081T-FF/FI, service manual, brief instructions and user manuals Emerson 5081T-FF/FI. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Emerson 5081T-FF/FI.

Similar manuals

A complete manual for the device Emerson 5081T-FF/FI, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Emerson 5081T-FF/FI by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Emerson 5081T-FF/FI.

A complete Emerson manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Emerson 5081T-FF/FI - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Emerson 5081T-FF/FI, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Emerson 5081T-FF/FI, that we can find in the current document
3. Tips how to use the basic functions of the device Emerson 5081T-FF/FI - which should help us in our first steps of using Emerson 5081T-FF/FI
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Emerson 5081T-FF/FI
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Emerson 5081T-FF/FI in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Emerson 5081T-FF/FI?

Use the form below

If you did not solve your problem by using a manual Emerson 5081T-FF/FI, ask a question using the form below. If a user had a similar problem with Emerson 5081T-FF/FI it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)