    Agilent T echnologies 54600-Series Oscilloscopes Data Sheet Multiple configurations to meet your needs If you w ork with bot h analog and digital components, Agilent T echnologies 54600-Series oscilloscopes can help you easil y see more of what's going on in your designs. The unique 2 + 16-channel mix ed signal oscillo- scope (MSO) models and ...

  • Agilent Technologies 54600-Series - page 2

    2 Selection Guide Figure 1. 2 + 16-channel mixed signal oscilloscope (MSO) family shown with 2- and 4-channel models. Selection Guide Model Bandwidth Maximum Sample Rate Maximum Memory* Channels 54621A 60 MHz 200 MSa/s 4 MB 2 54621D 60 MHz 200 MSa/s 4 MB 2+16 54622A 100 MHz 200 MSa/s 4 MB 2 54622D 100 MHz 200 MSa/s 4 MB 2+16 54624A 100 MHz 200 MSa/ ...

  • Agilent Technologies 54600-Series - page 3

    3 Figure 2. 2 + 16-channel mixed signal oscilloscope (MSO) allows you to view analog and digital content on one instrument. Mixed-signal scopes (models 54621D, 54622D, 54641D, and 54642D) The mix ed-signal oscilloscopes (MSO), with 2 analog channels and 16 digital channels, uniquely combine the det ailed signal analy- sis of a scope with t he multi ...

  • Agilent Technologies 54600-Series - page 4

    4 Affordable and Responsive MegaZoom deep memory With 4 MB t o 8 MB of MegaZoom deep memor y , 54600-Series scopes give y ou deep-memor y capture without t he sluggish response and complex operation you’v e had to t olerate wit h other deep-memor y scopes. And unlike the alter natives, MegaZoom deep memor y is not a special mode. That means it’ ...

  • Agilent Technologies 54600-Series - page 5

    5 Affordable and Responsive (continued) Revolutionary high-definition dis- play helps you see details Combine MegaZoom deep memor y with a patent ed high-def inition display system, and y ou will see subtle details t hat most scopes w on’t show y ou. The MegaZoom deep memor y is mapped into 32 lev els of int ensity on a display that has twice t h ...

  • Agilent Technologies 54600-Series - page 6

    6 Deep memory transfer over the bus Transfer all of y our dat a ov er t he bus, ev en when using deep memor y. T ypical transfer time is <70 seconds for 2 MB of data o ver GPIB. The 54640-Series requires soft- ware revision 1.0 or later . The 54620-Series requires software revision 1.5 or later . The syst em software is st ored in FL ASH ROM tha ...

  • Agilent Technologies 54600-Series - page 7

    7 N2758A CAN T rigger Module • T rigger on CAN 2.0A/B message frames with user specified IDs and/or data • T rigger on Error frames • Differential CAN • Acknowledge On/Off • Built In QuickHelp • Remote Control via GPIB/RS232 • Compatible with 54621D, 54622D, 54641D, 54642D The N2758A CAN trigger module extends your Mix ed Signal Oscil ...

  • Agilent Technologies 54600-Series - page 8

    8 T riggering Flexible T riggering With mixed analog and digital designs, sometimes it is hard to trace an anomaly back to its root cause unless you can trigger on it and correlat e it with anot her signal. With t he 54600-Series scopes, triggering is no longer a hassle. They come with f lexible triggering capabilities across all channels so you ca ...

  • Agilent Technologies 54600-Series - page 9

    9 T riggering (continued) I 2 C and SPI T riggering Don’t spend your time sorting through communication frames to find t he one of interest. Let Agilent set up an Int er -Int egrat ed Circuit (I 2 C) or Serial P eripheral Inter face (SPI) trigger to sif t through the frames for y ou. And like the C AN, LIN, and USB trig- gers, I 2 C and SPI come ...

  • Agilent Technologies 54600-Series - page 10

    10 MegaZoom deep memor y helps you determine ho w your signals are impacting each other . Wit h shallo w memor y scopes, you ha ve to choose whet her you look at a slo w analog signal or fast digit al content. With up t o 8 MB deep memor y , you don’t ha ve to choose – capture all of your data at once. Revolutionary high-def inition display re ...

  • Agilent Technologies 54600-Series - page 11

    11 Quickly pan and zoom f or analysis with MegaZoom’s inst ant response and optimum resolution. Autoscale lets y ou quickly display an y active signals, automatically setting the vertical, horizontal and trigger controls for the best displa y , while opti- mizing memor y. QuickMeas gives y ou any three automated measurements wit h the push of a b ...

  • Agilent Technologies 54600-Series - page 12

    12 Probes T o get t he most out of your scope, you need the right probes and accessories for y our par ticular application. That's why Agilent T echnologies of fers a com plet e family of inno vative probes and 54600-Series scopes. F or more comprehensiv e information, please see the A gilent 54600-Series Oscilloscopes Probes and A ccessories ...

  • Agilent Technologies 54600-Series - page 13

    13 Performance Characteristics Acquisition: Analog Channels Max Sample Rate 54621A/D, 54622A/D, 54624A: 200 MSa/s 54641A/D, 54642A/D: 2 GSa/sec interleaved, 1 GSa/sec each channel Max Memory Depth 54621A/D, 54622A/D, 54624A: 4 MB interleaved, 2 MB each channel 54641A/D, 54642A/D: 8 MB interleaved, 4 MB each channel V ertical Resolution 8 bits Peak ...

  • Agilent Technologies 54600-Series - page 14

    14 Performance Characteristics (continued) V ertical System: Analog Channels (continued) Range 1,2 54621A/D, 54622A/D, 54624A: 1 mV/div to 5 V/div 54641A/D, 54642A/D: 2 mV/div to 5 V/div Maximum Input CA T I 300 Vrms, 400 V pk, CA T II 100 Vrms, 400 Vpk With 10073C/10074C 10:1 probe: CA T I 500 Vpk, CA T II 400 Vpk 5 Vrms with 50 Ω input Offset R ...

  • Agilent Technologies 54600-Series - page 15

    15 Performance Characteristics (continued) V ertical System: Digital Channels (54621D, 54622D, 54641D, and 54642D only) Number of Channels 16 Digital – labeled D15 - D0 Threshold Groupings Pod 1: D7 - D0 Pod 2: D15 - D8 Threshold Selections TTL, CMOS, ECL, user-definable (selectable by pod) User-Defined Threshold Range ±8.0 V in 10 mV increments ...

  • Agilent Technologies 54600-Series - page 16

    16 Performance Characteristics (continued) Horizontal (continued) Digital Delta-t Accuracy 54621A/D, 54622A/D, 54624A: (non- V ernier settings) Same Channel: ±0.01% reading ±0.1% screen width ±(1 digital sample period, 2.5 or 5 ns based on sample rate of 200/400 MSa/s) Channel-to-Channel: ±0.01% reading ±0.1% screen width ±(1 digital sample p ...

  • Agilent Technologies 54600-Series - page 17

    17 Performance Characteristics (continued) T rigger System (continued) I 2 C T rigger on I 2 C (Inter-IC bus) serial protocol at a start/stop condition or user defined frame with address and/or data values. Also trigger on Missing Acknowledge, Restart, EEPROM read, and 10 bit write. SPI T rigger on SPI (Serial Protocol Interface) data pattern durin ...

  • Agilent Technologies 54600-Series - page 18

    18 Performance Characteristics (continued) Display System Display 7-inch raster monochrome CRT Throughput of Analog Channels 25 million vectors/sec per channel with 32 levels of intensity Resolution 255 vertical by 1000 horizontal points (waveform area) 32 levels of gray scale Controls W aveform intensity on front panel. Vectors on/off; infinite pe ...

  • Agilent Technologies 54600-Series - page 19

    19 Performance Characteristics (continued) Storage Save/Recall (non-volatile) 54621A/D, 54622A/D, 54624A: 3 setups and traces can be saved and recalled internally 54641A/D, 54642A/D: 4 setups and traces can be saved and recalled internally Floppy Disk 3.5” 1.44 MB double density Image formats : TIF , BMP Data formats: X and Y (time/voltage) value ...

  • Agilent Technologies 54600-Series - page 20

    20 Performance Characteristics (continued) Power Requirements Line V oltage Range 54621A/D, 54622A/D, 54624A: 100 - 240 V AC ±10%, CA T II, automatic selection 54641A/D, 54642A/D: 100-240 V AC, 50/60 Hz, CA T II, automatic selection; 100-132 VAC, 440 Hz, CA T II, automatic selection Line Frequency 54621A/D, 54622A/D, 54624A: 47 to 440 Hz 54641A/D, ...

  • Agilent Technologies 54600-Series - page 21

    21 Ordering Information Description 54621A 2-channel 60 MHz Oscilloscope 54621D 2+16 channel 60 MHz Mixed Signal Oscilloscope 54622A 2-channel 100 MHz Oscilloscope 54622D 2+16 channel 100 MHz Mixed Signal Oscilloscope 54624A 4-channel 100 MHz Oscilloscope 54641A 2-channel 350 MHz Oscilloscope 54641D 2+16 channel 350 MHz Mixed Signal Oscilloscope 54 ...

  • Agilent Technologies 54600-Series - page 22

    22 Available Options Description Option 003 Shielding Option for use in severe environments or with sensitive devices under test – shields both ways (in and out): RS-03 Magnetic interface shielding added to CRT , and RE-02 display shield added to CRT to reduce radiated interference Option 1CM Rackmount kit (same as 1186A) . W arranty and Calibrat ...

  • Agilent Technologies 54600-Series - page 23

    23 Ordering Information (continued) Fine Pitch Probing Description 10072A Fine-pitch probe kit 10075A 0.5 mm, IC clip kit E2613B 0.5mm, Wedge probe adapter , 3-signal, qty 2 E2614A 0.5 mm, Wedge probe adapter , 8-signal, qty 1 E2643A 0.5 mm, Wedge probe adapter 16-signal, qty 1 E2615B 0.65mm, Wedge probe adapter , 3-signal, qty 2 E2616A 0.65 mm, We ...

  • Agilent Technologies 54600-Series - page 24

    Call the measurement experts at Agilent T echnologies Whether your work is mostly digital, mostly analog, or somewhere in the middle, our measurement specialists can help you select the best debugging solution. Call today to talk to a knowledgeable engineer about your particular application. See for yourself with a no-obligation, 2-week demo If you ...

Производитель Agilent Technologies Категория Respiratory Product

Документы, которые мы получаем от производителя устройства Agilent Technologies 54600-Series мы можем разделить на несколько групп. Это в частности:
- технические чертежи Agilent Technologies
- инструкции обслуживания 54600-Series
- паспорта изделия Agilent Technologies
- информационные брошюры
- энергетические этикетки Agilent Technologies 54600-Series
Все из них важны, однако самую важную информацию с точки зрения пользователя мы найдем в инструкции обслуживания Agilent Technologies 54600-Series.

Группа документов, определяемая как инструкции обслуживания, делится также на более подробные типы, такие как: Инструкции монтажа Agilent Technologies 54600-Series, инструкции обслуживания, короткие инструкции или инструкции пользователя Agilent Technologies 54600-Series. В зависимости от потребностей, Вам необходимо поискать требуемый документ. На нашем сайте Вы можете просмотреть самую популярную инструкцию использования изделия Agilent Technologies 54600-Series.

Похожие инструкции обслуживания

Полная инструкция обслуживания устройства Agilent Technologies 54600-Series, как должна выглядеть?
Инструкция обслуживания, определяемая также как пособие пользователя, или просто "руководство" - это технический документ, цель которого заключается в использовании Agilent Technologies 54600-Series пользователями. Инструкции пишет, как правило технический писатель, языком, доступным для всех пользователей Agilent Technologies 54600-Series.

Полная инструкция обслуживания Agilent Technologies, должна заключать несколько основных элементов. Часть из них менее важная, как например: обложка / титульный лист или авторские страницы. Однако остальная часть, должна дать нам важную с точки зрения пользователя информацию.

1. Вступление и рекомендации, как пользоваться инструкцией Agilent Technologies 54600-Series - В начале каждой инструкции, необходимо найти указания, как пользоваться данным пособием. Здесь должна находится информация, касающаяся местонахождения содержания Agilent Technologies 54600-Series, FAQ и самых распространенных проблем - то есть мест, которые чаще всего ищут пользователи в каждой инструкции обслуживания
2. Содержание - индекс всех советов, касающихся Agilent Technologies 54600-Series, которое найдем в данном документе
3. Советы по использованию основных функций устройства Agilent Technologies 54600-Series - которые должны облегчить нам первые шаги во время использования Agilent Technologies 54600-Series
4. Troubleshooting - систематизированный ряд действия, который поможет нам диагностировать а в дальнейшем очередность решения важнейших проблем Agilent Technologies 54600-Series
5. FAQ - чаще всего задаваемые вопросы
6. Контактные данные Информация о том, где искать контактные данные производителя / сервисного центра Agilent Technologies 54600-Series в данной стране, если самостоятельно не получится решить проблему.

У вас вопрос, касающийся Agilent Technologies 54600-Series?

Воспользуйтесь формуляром, находящимся ниже

Если с помощью найденной инструкции Вы не решили свою проблему с Agilent Technologies 54600-Series, задайте вопрос, заполнив следующий формуляр. Если у какого то из пользователей была похожая проблема с Agilent Technologies 54600-Series со всей вероятностью он захочет поделиться методом ее решения.

Перепишите текст с картинки

Комментарии (0)