Manuale per l’uso Teledyne 9060Z

67 pagine 5.13 mb
Scarica

Vai alla pagina of 67

Summary
  • Teledyne 9060Z - page 1

     TELEDYNE  ANALYTICAL  INSTRUMENTS   Oxygen  Transmitter  Model  9060Z  Technical  Manual       ...

  • Teledyne 9060Z - page 2

    Model  9060Z  Manual   2     TABLE  OF  CONTENTS  TABLE  OF  CONTENTS.............. ...................................... .......................................................  1 1.  IMPORTANT  NOTICES................... ................. ................. ................. ..................... ...

  • Teledyne 9060Z - page 3

    Model  9060Z  Manual   3    5.3  KEYPAD.............. ...................................... ................... ................ ................. ................  28 5.3.1  Keypad  in  RUN  mode  ........................................................................... ...............  29 5.3.2 ? ...

  • Teledyne 9060Z - page 4

    Model  9060Z  Manual   4    8.1  FUNCTION  SUMMARY  TABLE  ................ ...................................... ................................  42 8.2  CALIBRATION  MODE  FUNCTIONS  ................ ..................... ............... ............................  43 8.2.1  Reference ? ...

  • Teledyne 9060Z - page 5

    Model  9060Z  Manual   5     Copyright  TELEDYNE  ANALYTICAL  INSTRUMENTS  —  2012  This  manual  describes  the  transmitter  firmware  version  1.24,  September  2012  Neither  the  whole  nor  any  part  of  the  information  contained  in, ...

  • Teledyne 9060Z - page 6

    Model  9060Z  Manual   6                             Important  Notice  Regarding  9060 Probe  Option ‐ FI L ‐ 3   WARNING:  The  only  identifiable  standard  for  flame  arresters  for  gener ...

  • Teledyne 9060Z - page 7

    Model  9060Z  Manual   7     1. IMPORTANT  NOTICES  This  manual  is  intended  as  a  supplement  to  the  9060Z  Operators  Manual.  It  is  intended  to  be  used  by  technical  personal  that  are  qualified  to  install,  commission, ...

  • Teledyne 9060Z - page 8

    Model  9060Z  Manual   8    2.  INTRODUCTION  2.1  9060Z  TRANSMITTER  The  Teledyne  Analytical  Instruments  9060Z  oxygen  transmitter  is  designed  to  measure  the  oxygen  gas  concentration  in  combustion  applications  such  as  power  st ...

  • Teledyne 9060Z - page 9

    Model  9060Z  Manual   9    Teledyne  Analytical  Instruments  series  9060H  oxygen  probes  and  sensors  employ  state ‐ of ‐ the ‐ art  zirconia  sensors  and  advanced  materials,  which  provide  the  following  benefits:   Improved  control  ...

  • Teledyne 9060Z - page 10

    Model  9060Z  Manual   10    Caution,  risk  of  danger.  See  additional  information  in  the  manual.  3.1  HARDWARE  SPECIFICATIONS  3.1.1  TRANSMITTER  SPECIFICATIONS  Number  of  oxygen  probes:  2  maximum  Oxygen  range:  0 ‐ 1%  to ? ...

  • Teledyne 9060Z - page 11

    Model  9060Z  Manual   11    Probe  insertion  length  (from  process  end  of  mounting  thread  to  probe  sensing  tip.  3.1.3  SERIES  9060H  PROBES  SPECIFICATIONS  MODEL  9060H 9060UL/UH Application  Combustion  flue  Combustion  flue  gases ...

  • Teledyne 9060Z - page 12

    Model  9060Z  Manual   12     3.1.4  OXYGEN  PROBE  MODEL  SELECTION  GUIDE  Heated  probes ‐ temperature  range  0 ‐ 900°C  (1650°F).     *Note:  (1)  The  Inconel  option  has  all  inconel  wetted  parts  except  for  the  ceramic ? ...

  • Teledyne 9060Z - page 13

    Model  9060Z  Manual   13      Unheated  probes  for  clean  gases ‐ temper atur e  range  700 ‐ 1400°C  (1290 ‐ 2550°F).      *Note:  (2)  A  standard  oxygen  probe  for  carburising  furnaces,  has  a  253  MA  sheath.     ...

  • Teledyne 9060Z - page 14

    Model  9060Z  Manual   14     3.2  OPERATIONAL  SPECIFICATIONS  Range  of  outputs:  Function    Minimum Maximum  Linear  oxygen  #1    0  to  1%    0  to  100%  (**)  Very  low  oxygen  #1    0  to  0.0 ...

  • Teledyne 9060Z - page 15

    Model  9060Z  Manual   15    The  oxygen  deficiency  output  can  be  used  in  the  same  way  as  a  combustibles  analyzer  to  signal  the  extent  of  reducing  conditions  of  combustion  processes.  4.  INSTALLATION  AND  COMMISSIONIN ...

  • Teledyne 9060Z - page 16

    Model  9060Z  Manual   16    Probes  can  be  mounted  at  any  angle.  However,  if  the  probe  is  to  be  mounted  on  a  vertical  duct  wall,  it  is  better  to  angle  the  probe  (approx.  15°)  down  to  avoid  process ...

  • Teledyne 9060Z - page 17

    Model  9060Z  Manual   17    4.3  INSTALLING  THE  AUXILIARY  THERMOCOUPLE  Weld  a  1/2  inch  BSP  mounting  socket  to  the  flue  within  about  300  mm  (12”),  and  upstream  of  the  oxygen  probe.  The  thermocouple  should  be ? ...

  • Teledyne 9060Z - page 18

    Model  9060Z  Manual   18    Connection  Diagram  for  9060Z  Transmitter  and  one  or  two  9060H  Heated  Sensors  ...

  • Teledyne 9060Z - page 19

    Model  9060Z  Manual   19    Connection  Diagram  for  9060Z  Transmitter  and  one  or  two  9060UL/UH  Unheated  Probes   ...

  • Teledyne 9060Z - page 20

    Model  9060Z  Manual   20    • The  mains  power  must  be  either  100/110VAC  or  220/240VAC  with  a  mains  frequency  of  50  or  60Hz  • The  supply  circuit  must  be  fused  to  at  least  10Amps  and  have  a  dual  pole ...

  • Teledyne 9060Z - page 21

    Model  9060Z  Manual   21     Connection  of  Probe  Cable  for  Unheated  Probes  Models  9060UL/UH.   Connection  of  Probe  Cable  for  Heated  Probes  Model  9060H  .   ...

  • Teledyne 9060Z - page 22

    Model  9060Z  Manual   22    4.8  CONNECTING  THE  AUXILIARY  THER MOCOUPLE  (OPTIONAL)  For  9060H  heated  probes,  the  auxiliary  thermocouple  must  be  a  separate  TC  with  the  junction  isolated  from  earth,  mounted  near  to  and ? ...

  • Teledyne 9060Z - page 23

    Model  9060Z  Manual   23    The  on ‐ line  auto  purge  and  calibration  check  system  is  optional.  For  details  on  its  operation  see  chapters  4.24,  4.25  and  chapter  11.  To  automatically  sense  a  blocked  probe  filter, ...

  • Teledyne 9060Z - page 24

    Model  9060Z  Manual   24    network  line  from  the  transmitters  is  taken  from  the  middle  of  the  transmitter  network  string,  a  terminating  resistor  should  be  enabled  with  LK3  at  each  end  of  the  network  line.  ...

  • Teledyne 9060Z - page 25

    Model  9060Z  Manual   25    high  impedance  alarm,  ‘Probe  1  (2)  High  Impedance’,  is  not  active.  The  actual  impedance  can  be  displayed  on  the  lower  line.  It  should  be  less  than  9  K Ω at  720°C  (1320°F). ...

  • Teledyne 9060Z - page 26

    Model  9060Z  Manual   26    downwards  with  the  filter  removed.  Probes  can  also  be  mounted  horizontally  with  no  filter  with  some  dusts.  An  occasional  automatic  back  purge  is  helpful  in  this  case.  Normally  heated ? ...

  • Teledyne 9060Z - page 27

    Model  9060Z  Manual   27    5.  DISPLAY  AND  KEYPAD  Operator  interaction  with  the  oxygen  transmitter  is  done  through  a  graphical  display  and  8  keys  on  the  front  of  the  case.  The  five  LED  indicators  ar e  used ...

  • Teledyne 9060Z - page 28

    Model  9060Z  Manual   28     5.2  OXYGEN  DISPLAY  UNITS  The  measured  oxygen  content  displayed  on  the  top  line  of  the  transmitter  for  probe  1  and  on  the  lower  line  for  probe  2  can  be  shown  in  either ? ...

  • Teledyne 9060Z - page 29

    Model  9060Z  Manual   29    SETUP  /  RUN  key  Pressing  this  key  will  put  the  transmitter  into  one  of  the  menu ‐ driven  SETUP,  C OMMISSIONING  or  CALIBRATION  modes  (see  chapters  6,  7  and  8  for  details).  The  ...

  • Teledyne 9060Z - page 30

    Model  9060Z  Manual   30    FUNCTION  up  and  FUNCTION  down  keys  These  two  keys  allow  for  the  selection  of  the  setup  function.  A  function  summary  table  is  found  at  the  start  of  the  relevant  SETUP,  COMMISSION ...

  • Teledyne 9060Z - page 31

    Model  9060Z  Manual   31    6.  SETUP  MODE  The  SETUP  mode  is  mode  is  accessed  by  pressing  the  SETUP  key  momentarily  so  the  words  ‘Setup  Mode’  appear  at  the  bottom  of  the  display.  The  transmitter  will ? ...

  • Teledyne 9060Z - page 32

    Model  9060Z  Manual   32    6.3  SETUP  MODE  FUNCTIONS   6.3.1  Probe  1  Offset  Each  Teledyne  Analytical  Instruments  probe  has  an  offset  calibration  value  printed  on  a  tag  that  is  attached  to  the  probe  when  it ? ...

  • Teledyne 9060Z - page 33

    Model  9060Z  Manual   33      NOTE:  An  asterisk  (*)  on  the  end  of  the  line  identifies  the  ite m  is  enabled  by  default  after  a  COLD ‐ START.  6.3.3  Oxygen  Display  Units  The  top  line  of  the  display  th ...

  • Teledyne 9060Z - page 34

    Model  9060Z  Manual   34    7.  COMMISSIONING  MODE  This  chapter  describes  the  functions  available  when  the  COMMISSIONING  mode  is  entered  on  the  transmitter.  The  COMMISSIONING  mode  is  accessed  by  pressing  and  holding  th ...

  • Teledyne 9060Z - page 35

    Model  9060Z  Manual   35      ...

  • Teledyne 9060Z - page 36

    Model  9060Z  Manual   36    7.2  COMMISSIONING  MODE  FUNCTIONS  7.2.1  Internal  Date  /  Time  The  date  and  time  are  used  in  the  transmitter  to  run  time  based  ope rat ions  such  as  impedance  readings  and  gas  calibrati ...

  • Teledyne 9060Z - page 37

    Model  9060Z  Manual   37     The  zero  and  span  of  the  selected  output  are  set  in  the  following  two  menus  (functions  11&12  and  14&15).  NOTE:  Regarding  channel  output  configuration;  both  4 ‐ 20mA  channels  a ...

  • Teledyne 9060Z - page 38

    Model  9060Z  Manual   38    7.2.8  Flue  Pressure  Units  and  Value  The  9060Z  transmitter  is  capable  of  fixed  or  variable  input  pressure  compensation  in  the  calculation  of  oxygen  from ‐ 1  Atm  to  +3  Atm.  If  th ...

  • Teledyne 9060Z - page 39

    Model  9060Z  Manual   39    7.2.16  Solenoid  1  &  2  Post  Freeze  Duration  of  time  the  transmitter  waits  before  resuming  live  readings  following  a  gas  calibration  or  purge  event.  (See  also  chapter  7.2.10)  7.2.17 ? ...

  • Teledyne 9060Z - page 40

    Model  9060Z  Manual   40    All  of  the  items  in  this  list  are  selected  as  the  default  setting.  Any  of  these  items  can  be  disabled  from  the  common  alarm  by  pressing  the  ENTER  key,  and  they  will  then ...

  • Teledyne 9060Z - page 41

    Model  9060Z  Manual   41    7.2.25  Communications  Port  Options  The  9060Z  transmitter  has  a  serial  communications  port  available  at  terminals  18  to  22.  The  defaul t  protocol  is  for  RS ‐ 232  running  at  19,200  baud  ...

  • Teledyne 9060Z - page 42

    Model  9060Z  Manual   42    8.  CALIBRATION  MODE  This  chapter  describes  the  functions  available  when  the  CALIBRATION  mode  is  selected  on  the  transmitter.  For  specific  information  about  calibrating  the  transmitter  see  chapt ...

  • Teledyne 9060Z - page 43

    Model  9060Z  Manual   43    8.2  CALIBRATION  MODE  FUNCTIONS  8.2.1  REFERENCE  VOLTAGES  The  calibration  of  the  analogue  inputs  is  based  on  the  voltage  of  a  temper ature  compensated  voltage  reference  integrated  circuit.  There ...

  • Teledyne 9060Z - page 44

    Model  9060Z  Manual   44    8.2.8  HEATER  SSR  FAULT  CORRECTION  In  the  event  of  one  of  the  heater  solid  state  relays  (SSR)  failing  in  single  probe  mode,  the  heater  dri ve  output  can  be  swapped  to  the  HEA ...

  • Teledyne 9060Z - page 45

    Model  9060Z  Manual   45    9.  ALARMS  The  9060Z  has  4  alarm  relays  and  a  built  in  alarm  annunciator.  If  an  alarm  occurs,  the  ALARM  light  will  flash.  To  find  out  what  the  alarm  is,  press  the  ALARM ? ...

  • Teledyne 9060Z - page 46

    Model  9060Z  Manual   46    4.  ‘Probe  2  High  Impedance’  Oxygen  probe  or  electrode  failure.  The  sensor  has  a  high  impedance  measurement.  Replac e  the  probe.  This  alarm  is  inhibited  when  the  probes  are  less  ...

  • Teledyne 9060Z - page 47

    Model  9060Z  Manual   47    19.  ‘Gas  1  Calibration  Error’  20.  ‘Gas  2  Calibration  Error’  At  the  end  of  an  automatic  gas  calibration  check  cycle  the  oxygen  reading  was  outside  the  values  set  in  the  C ...

  • Teledyne 9060Z - page 48

    Model  9060Z  Manual   48    level.  With  heated  probes  this  relay  will  be  de ‐ energised  while  the  probe  is  heating  up  from  ambient,  making  the  contacts  open  circuit.  30.  ‘Cal  1  in  Progress’  31.  ‘Cal  2 ...

  • Teledyne 9060Z - page 49

    Model  9060Z  Manual   49    10.  INSTRUMENT  CALIBRATION  10.1  CALIBRATION  SUMMARY  The  9060Z  oxygen  transmitter  has  a  self  calibration  and  diagnostic  system  built  into  the  hardware  and  software.  Once  the  reference  voltages ? ...

  • Teledyne 9060Z - page 50

    Model  9060Z  Manual   50    11. The  output  should  go  20. 00mA.  The  value  in  function  #7  (10)  can  be  fine  tuned  if  necessary.  12. Go  back  to  function  #5  (8 )  and  select  ‘Calibrate  4mA’  again  and  check  t ...

  • Teledyne 9060Z - page 51

    Model  9060Z  Manual   51    11.  GAS  CALIBRATION  CHECK  AND  PURGE  The  oxygen  sensor  that  is  used  in  the  Teledyne  Analytical  Instruments  oxygen  probe  is  extrem ely  predictable,  stable  and  reliable.  For  this  reason,  t ...

  • Teledyne 9060Z - page 52

    Model  9060Z  Manual   52     12.  SOFTWARE  UPGRADES  The  software  for  the  9060Z  series  transmitters  is  contained  in  memory  inside  the  microprocessor  integrated  circuit  (IC).  The  transmitter  therefore  does  not  require  an ...

  • Teledyne 9060Z - page 53

    Model  9060Z  Manual   53    13.  TROUBLESHOOTING  The  Teledyne  Analytical  Instruments  oxygen  transmitter  has  proved  an  extremely  reliable  instru ment.  The  trouble  shooting  guide  here  is  ba sed  on  an  analysis  of  the  potentia ...

  • Teledyne 9060Z - page 54

    Model  9060Z  Manual   54    14.  MAINTENANCE  14.1  TRANSMITTER  MAINTENANCE  The  9060Z  transmitter  has  several  hardware  checking  systems  that  confirm  the  correct  operation  of  the  transmitter  and  raise  an  alarm  if  there  is ...

  • Teledyne 9060Z - page 55

    Model  9060Z  Manual   55    15.  INDEX  9060H  Oxygen  Probes  .................................... .................................... ................ ............... ................  9  Alarm  connections  ................ .................................... .................. .............. ...

  • Teledyne 9060Z - page 56

    Model  9060Z  Manual   56    Probe  offset  entry  .................. .................. .................. .................. ................. ............. ............ ..........  34  Purge  ................. ................ ................................ ................................ ............ ...

  • Teledyne 9060Z - page 57

    Model  9060Z  Manual   57    APPENDIX  1,  CONSTITUENT  VALUES  FOR  VARIOUS  FUELS  If  the  transmitter  is  set  up  to  have  readout  or  output  of  combustibles,  oxygen  deficiency,  maximum  carbon  dioxide  or  combustion  efficiency, ...

  • Teledyne 9060Z - page 58

    Model  9060Z  Manual   58    APPENDIX  2,  PROBE  EMF  TABLES  ZIRCONIA  OXYGEN  SENSOR  OUTPUT  (mV)  PROBE  TYPE  9060H,  SENSOR  OXYGEN  %  Probe  EMF  @  720°C  (1320°F)  OXYGEN  %  Probe  EMF  @  720°C  (1320°F)  20 ...

  • Teledyne 9060Z - page 59

    Model  9060Z  Manual   59     ZIRCONIA  OXYGEN  PROBE  OUTPUT  (mV)  PROBE  TYPE  9060UL/UH   These  tables  are  based  on  the  Nernst  equation:  ...

  • Teledyne 9060Z - page 60

    Model  9060Z  Manual   60    EMF  =       ₂ .  .  Where  T  is  temperature°  K  Thermocouple  information  from  NIST  ITS ‐ 90  data  tables  APPENDIX  3,  %  OXYGEN  SCALE  TO  LOGARITHMIC  OXYGEN ? ...

  • Teledyne 9060Z - page 61

    Model  9060Z  Manual   61      APPENDIX  4,  MODBUS™  The  Teledyne  Analytical  Instruments  9060Z  transmitter  has  the  ability  to  work  as  a  Modbus  slave  node  on  serial  RS ‐ 232  or  3 ‐ wire  RS ‐ 485  via  RTU  mode ...

  • Teledyne 9060Z - page 62

    Model  9060Z  Manual   62    Alarm  status  is  stored  as  an  integer  value:  0  clear  2  active  1  self  cleared  3  acknowledged  9060Z  Transmitter  Modbus  Input  Register  Table ...

  • Teledyne 9060Z - page 63

    Model  9060Z  Manual   63       ...

  • Teledyne 9060Z - page 64

    Model  9060Z  Manual   64     ...

  • Teledyne 9060Z - page 65

    Model  9060Z  Manual   65    9060Z  Transmitter  Modbus  Holding  Register  Table  for  firmware  v1.16  and  newer  ...

  • Teledyne 9060Z - page 66

    Model  9060Z  Manual   66          ...

  • Teledyne 9060Z - page 67

    Model  9060Z  Manual   67       0x41  Special  Instruction  Function  This  command  allows  you  to  interact  with  the  transmitter  allo wing  access  to  functionality  that  would  otherwise  only  be  accessible  via  the  keypad.  ...

Fabbricante Teledyne Categoria Oxygen Equipment

I documenti che otteniamo dal fabbricante del dispositivoTeledyne 9060Z possono essere suddivisi in diversi gruppi. Questi sono tra gli altri:
- disegni tecnici Teledyne
- manuali per l’uso 9060Z
- schede prodotto Teledyne
- opuscoli
- o etichette energetiche Teledyne 9060Z
Tutti sono importanti, ma le informazioni più importanti dal punto di vista di utilizzo del dispositivo possono essere trovate nel manuale per l’uso Teledyne 9060Z.

Un gruppo di documenti denominato manuali per l’uso, è anche suddiviso in tipi più specifici, come: Manuali di installazione Teledyne 9060Z, manuali di manutenzione, brevi manuali o manuali utente Teledyne 9060Z. A seconda delle esigenze, si dovrebbe cercare il documento necessario. Nel nostro sito web, puoi visualizzare il manuale più popolare per l'uso del prodotto Teledyne 9060Z.

Manuali per l’uso simili

Manuale completo del dispositivo Teledyne 9060Z, come deve essere?
Il manuale per l’uso, denominato anche istruzioni per l’uso, o semplicemente il manuale è un documento tecnico progettato per aiutare ad utilizzare Teledyne 9060Z dagli utenti. I manuali sono di solito scritti da uno scrittore tecnico, ma in un linguaggio accessibile a tutti gli utenti Teledyne 9060Z.

Il manuale per l’uso completo Teledyne, dovrebbe includere vari elementi di base. Alcuni di loro sono meno importanti, come ad esempio: copertina / pagina del titolo o pagina di autore. Tuttavia, la parte rimanente, dovrebbe fornire informazioni importanti dal punto di vista dell'utente.

1. Introduzione e la guida su come utilizzare il manuale Teledyne 9060Z - All'inizio di ogni manuale, dovremmo trovare indizi su come utilizzare il documento. Dovrebbe contenere informazioni sulla posizione dell’indice Teledyne 9060Z, FAQ o i problemi più comuni : i punti che sono più spesso cercati dagli utenti di ogni manuale
2. Indice - elenco di tutti i suggerimenti per Teledyne 9060Z che possono essere trovati nel documento corrente
3. Suggerimenti su come utilizzare le funzioni di base del dispositivo Teledyne 9060Z - che ci dovrebbe aiutare nei primi passi durante l'uso Teledyne 9060Z
4. Troubleshooting - sequenza sistematica delle attività che vi aiuterà a diagnosticare e, successivamente, risolvere i problemi più importanti del Teledyne 9060Z
5. FAQ - domande più frequenti
6. Dettagli di contatto Informazioni su dove cercare il contatto con il fabbricante / servizio Teledyne 9060Z in un dato paese, se il problema non può essere risolto da noi stessi.

Hai una domanda su Teledyne 9060Z?

Utilizza il modulo sottostante

Se non hai risolto il problema con Teledyne 9060Z con l'aiuto del manuale che hai trovato, fai una domanda utilizzando il modulo sottostante. Se un utente ha avuto un problema simile con Teledyne 9060Z è probabile che voglia condividere il modo di risolverlo.

Riscrivi il testo dall’immagine

Commenti (0)