    70/73 English Instru ction Sheet Page 1 ® 70/73 Series III Multimeter Instruction Sheet W Read First: Safety Information To ensure that the meter is us ed safely, fo llow these i nstructions: • Never use the m eter if the mete r or test lea ds look damaged. • Be sure the tes t leads and rot ary switc h are in the correc t position for t he des ...

  • Fluke PN 650454 - page 2

    70/73 English Instru ction Sheet Page 2 Input Jacks 300 mA 10A COM FUSED V ! Milliamps Amps Volts, Ohms, Diode Test Common Terminal 600V 70 70 hr2f.eps See Specific ations for overl oad protection. Pushbutton Use the pushb utton to eith er select a fix ed range or p ut the meter in the Automat ic Touch Hold  mode. Autorange The meter defa ults t ...

  • Fluke PN 650454 - page 3

    70/73 English Instru ction Sheet Page 3 Automatic Touch Hold  Mode W Warning To avoid electric shock, do not use the Touch Hold  to determine if a circ uit with high volt age is dead. The Touch Hold  mode will not capture unstable or noisy readings. The Touch H old  mode autom atically captures and d isplays stable readi ngs. Press R fo ...

  • Fluke PN 650454 - page 4

    70/73 English Instru ction Sheet Page 4 Standby If the meter is on but is inac tive for an ho ur (20 minutes i n diode test), the sc reen goes blank and display s selected ba r graph segments . To resume operat ion, turn the r otary switch or pr ess R . AC and DC Voltage ( K L m L ) + _ + _ V AC V DC + _ Volts AC Volts DC Millivolts DC V V V mV hr4 ...

  • Fluke PN 650454 - page 5

    70/73 English Instru ction Sheet Page 5 Diode Test ( G ) Forward Bias Single Beep Reverse Bias Shorted Open or Good Diode Bad Diode + _ + _ Bad Diode + _ Good Diode + _ hr7f.eps ...

  • Fluke PN 650454 - page 6

    70/73 English Instru ction Sheet Page 6 Continuity Test ( R ) + _ + _ hr5f.eps If continuity exists (res istance < 210 Ω ), the beeper s ounds continuous ly. The met er beeps twice if it is in th e Touch Hold  mode. Current ( ? A ) W Warning To avoid injury, do n ot attempt a current measurement if the open circuit voltage is above the rate ...

  • Fluke PN 650454 - page 7

    70/73 English Instru ction Sheet Page 7 Holster The snap-on h olster absorbs sho cks and protects the meter an d comes with a Flex -Stand  . To pro tect the front of the meter, put the meter fac e down in the holster. hr15f.eps Maintenanc e W Warning To avoid electr ic shock, remo ve the test lead s before opening t he case, and clos e the case ...

  • Fluke PN 650454 - page 8

    70/73 English Instru ction Sheet Page 8 Internal Fuse Test 300 mA Ω 5-8 Ω <0.5 OK OK 10 A OK 70 OK hr10f.eps Battery and Fuse Re placeme nt Note Before opening the case, mak e sure the tes t leads are removed and the rotary sw itch is turned to OFF. F1 F2 70 hr11f.eps ...

  • Fluke PN 650454 - page 9

    70/73 English Instru ction Sheet Page 9 Cleaning To clean the m eter, use a dam p cloth and m ild detergent; d o not use abrasiv es or solvents o n the meter. Service a nd Part s For servic e information in t he U.S.A. and C anada, call 1-8 00-44- FLUKE (1-800-44 3-5853), in Europe, 31 40 267 8200, from other countries, 1- 425-356-5500 . Item Descr ...

  • Fluke PN 650454 - page 10

    70/73 English Instru ction Sheet Page 10 Electromagnet ic Compatibility in RF field of 3 V/m on all functi ons exc ept M Ω Total a ccurac y = Spe cified ac curacy plus + 2.0% of range for all functi ons excep t: 320.0 mV range: tot al acc uracy = specified accuracy + 1.0% of range 320.0 Ω range: total accuracy = specifi ed accuracy + 8% of rang ...

  • Fluke PN 650454 - page 11

    70/73 English Instru ction Sheet Page 11 Function Range Accuracy 0L 3.200 V, 32.00 V, 320.0 V 600 V ± (0.3%+1) ± (0.4%+1) m L 320.0 mV ± (0.3%+1) K (45 to 500 Hz, 3.2V range. Other ranges 45 to 1 kHz) 3.200 V, 32.00 V, 320.0 V, 600 V ± (2%+2) ± (2%+2) 320.0 Ω 3200 Ω , 32.00 k Ω , 320.0 k Ω , 3.200 M Ω 32.00 M Ω ± (0.5%+2) ± (0.5% ...

  • Fluke PN 650454 - page 12

    70/73 English Instru ction Sheet Page 12 Overload pro tection for all functions and r anges: 600 V. Function Input Impedance (Nominal) L , m L , K >10 M Ω , <50 pF Common Mo de Rejection Rat io (1 k Ω Un balanced ) Normal Mode Rejection L , m L >120 dB at dc, 50 Hz, or 60 Hz >60 dB at 50 Hz or 60 Hz K >60 dB dc to 60 Hz Open Circ ...

Fabbricante Fluke Categoria Impact Driver

I documenti che otteniamo dal fabbricante del dispositivoFluke PN 650454 possono essere suddivisi in diversi gruppi. Questi sono tra gli altri:
- disegni tecnici Fluke
- manuali per l’uso PN 650454
- schede prodotto Fluke
- opuscoli
- o etichette energetiche Fluke PN 650454
Tutti sono importanti, ma le informazioni più importanti dal punto di vista di utilizzo del dispositivo possono essere trovate nel manuale per l’uso Fluke PN 650454.

Un gruppo di documenti denominato manuali per l’uso, è anche suddiviso in tipi più specifici, come: Manuali di installazione Fluke PN 650454, manuali di manutenzione, brevi manuali o manuali utente Fluke PN 650454. A seconda delle esigenze, si dovrebbe cercare il documento necessario. Nel nostro sito web, puoi visualizzare il manuale più popolare per l'uso del prodotto Fluke PN 650454.

Manuali per l’uso simili

Manuale completo del dispositivo Fluke PN 650454, come deve essere?
Il manuale per l’uso, denominato anche istruzioni per l’uso, o semplicemente il manuale è un documento tecnico progettato per aiutare ad utilizzare Fluke PN 650454 dagli utenti. I manuali sono di solito scritti da uno scrittore tecnico, ma in un linguaggio accessibile a tutti gli utenti Fluke PN 650454.

Il manuale per l’uso completo Fluke, dovrebbe includere vari elementi di base. Alcuni di loro sono meno importanti, come ad esempio: copertina / pagina del titolo o pagina di autore. Tuttavia, la parte rimanente, dovrebbe fornire informazioni importanti dal punto di vista dell'utente.

1. Introduzione e la guida su come utilizzare il manuale Fluke PN 650454 - All'inizio di ogni manuale, dovremmo trovare indizi su come utilizzare il documento. Dovrebbe contenere informazioni sulla posizione dell’indice Fluke PN 650454, FAQ o i problemi più comuni : i punti che sono più spesso cercati dagli utenti di ogni manuale
2. Indice - elenco di tutti i suggerimenti per Fluke PN 650454 che possono essere trovati nel documento corrente
3. Suggerimenti su come utilizzare le funzioni di base del dispositivo Fluke PN 650454 - che ci dovrebbe aiutare nei primi passi durante l'uso Fluke PN 650454
4. Troubleshooting - sequenza sistematica delle attività che vi aiuterà a diagnosticare e, successivamente, risolvere i problemi più importanti del Fluke PN 650454
5. FAQ - domande più frequenti
6. Dettagli di contatto Informazioni su dove cercare il contatto con il fabbricante / servizio Fluke PN 650454 in un dato paese, se il problema non può essere risolto da noi stessi.

Hai una domanda su Fluke PN 650454?

Utilizza il modulo sottostante

Se non hai risolto il problema con Fluke PN 650454 con l'aiuto del manuale che hai trovato, fai una domanda utilizzando il modulo sottostante. Se un utente ha avuto un problema simile con Fluke PN 650454 è probabile che voglia condividere il modo di risolverlo.

Riscrivi il testo dall’immagine

Commenti (0)