    Ultraviolet Crosslinkers Operating Instructions and Service Manual IMPORTANT: Please read these instructions before operating your UVP UV Crosslinker to familiarize yourself with its operation. 81-0112-01 Rev C ...

  • UVPS 81-0112-01 Rev C - page 2

    UV CROSSLINKERS UVP CONTENTS 1. Introduction 1 2. Important Safety Information 1 3. Description and Specifications 1 4. Operation 3 5. Applications 6 6. Maintenance, Care and Cleaning 7 7. Changing the UV Wavelength 8 8. Calibration Procedure 10 9. Replacement Parts 12 10. General Servicing Procedures 13 Schematic Diagram 19 ...

  • UVPS 81-0112-01 Rev C - page 3

    UV CROSSLINKERS UVP 1. INTRODUCTION UVP Crosslinkers offer researchers an instrume nt to quickly, safely and efficiently expose samples to a controlled amount of ultraviolet radiation. CL-1000 series features a lay-down type drawer while the CX-2000 has a pull-out dr awer. Exposure of samples to UV provides for the following: • Crosslinking of DN ...

  • UVPS 81-0112-01 Rev C - page 4

    UV CROSSLINKERS UVP Features • Microprocessor controlled/UV sensor feedback system • Multiple set functions- preset UV energy exposure preset UV time exposure user set UV energy exposure user set UV time exposure • Maximum UV energy exposure setting of 999,900 m icrojoules per cm2 • Maximum UV tim e exposure setting of 999.9 minutes • Int ...

  • UVPS 81-0112-01 Rev C - page 5

    UV CROSSLINKERS UVP 4. OPERATION 1. Place the crosslinker on a level work surface. Be sure there is enough room in front for the door to open easily. 2. Plug the female end of the power cord into the crosslinker. 3. Plug the male end of the power cord into a properly grounded electrical outlet. The proper operating voltage of the crosslinker is fou ...

  • UVPS 81-0112-01 Rev C - page 6

    UV CROSSLINKERS UVP The new setting will appear on the LED display. Push ENTER on the touchpad. The new setting is now installed. B. Preset ultraviolet time exposure Push PRESET and then TIME on the tactile touchpad. The red light at each position should now be lit and the prepr ogramm ed UV exposure time of 2 minutes should be displayed in the LED ...

  • UVPS 81-0112-01 Rev C - page 7

    UV CROSSLINKERS UVP Push START on the touchpad. After a slight delay to energize the UV tubes, The LED will begin to countdown. the unit will stop automatically at the end of the exposure cycle and will beep five times. Exposure is now complete. D. User set UV time exposure Setting your own UV time exposure can be achieved as follows: Push the TIME ...

  • UVPS 81-0112-01 Rev C - page 8

    UV CROSSLINKERS UVP 5. APPLICATIONS The Crosslinker is a multi-purpose ultraviolet e xposure instrum ent for use in the laboratory. A wide variety of applications for ultraviolet radiation exist in the laboratory. A. UV crosslinking of DNA and RNA by covale ntly binding nucleic acids to transfer membranes nitrocellulose, nylon or nylon-reinforced n ...

  • UVPS 81-0112-01 Rev C - page 9

    UV CROSSLINKERS UVP 6. MAINTENANCE, CARE AND CLEANING OF THE CROSSLINKER The UVP UV Crosslinker is built to provide you w ith trouble-free operation. To ensure correct service: NOTE: ALWAYS UNPLUG UNIT FROM ELEC TRICAL SUPPLY BEFORE CLEANING OR DRYING. 1. Wipe ANY water from inside and outside the unit with a soft cloth or sponge. 2. Use soap and w ...

  • UVPS 81-0112-01 Rev C - page 10

    UV CROSSLINKERS UVP 7. CHANGING THE UV WAVELENGTH A UV Crosslinker is purchased with shortwave, l ongwave or midrange tubes. However, if user requirements and applications change, the Crosslinkers provide you with the unique ability to change the ultraviolet wavelength and recalibrate the UV sensor and microprocessor to the new UV wavelength. This ...

  • UVPS 81-0112-01 Rev C - page 11

    UV CROSSLINKERS UVP 11. The previous calibrated operation of the crosslinker will continue following any unsuccessful recalibration. Replace the new UV wavelength tubes with the previously removed old wavelength tubes. 12. Successful recalibration to the NEW UV wavelength changes the values in the microprocessor and numeric setting value is display ...

  • UVPS 81-0112-01 Rev C - page 12

    UV CROSSLINKERS UVP 8. CROSSLINKER CALIBRATION PROCEDURE Equipment necessary: Chart recorder Two probes with banana plug adapters One radiometer with UVX-25 calibrated sensor Cable with banana plug and jack adaptor to connect radiometer to chart-recorder Power source (100V, 115V or 230V) Procedure: 1. Connect the crosslinker to the power source. Do ...

  • UVPS 81-0112-01 Rev C - page 13

    UV CROSSLINKERS UVP 9. Run the Crosslinker in calibration mode by depressing the STOP/RESET button and turning the power switch on at the same time. At this point, both pens should start recording information on paper. Som etimes either pen will record zero inform ation regardless of the lights in the crosslinker be ing on, if this is the case, che ...

  • UVPS 81-0112-01 Rev C - page 14

    UV CROSSLINKERS UVP 9. REPLACEMENT PARTS AND SERVICING PROCEDURES PART NO. A. OPERATIONS WITHOUT REMOVING THE TOP COVER 5 x 34-0007-01 Changing tubes B. OPERATIONS WITH TOP COVER REMOVED Replacement of tube holders 5 x 36-0003-01 Replacement of starters 5 x 53-0001-02 or 53-0132-01 Replacement of starter holders 5 x 36-0005-01 Replacement of fuse h ...

  • UVPS 81-0112-01 Rev C - page 15

    UV CROSSLINKERS UVP 10. GENERAL SERVICING PROCEDURES A. Operations Without Removing The Top Cover Changing the Tubes It is recommended to change all FIVE 8-W att 254nm UV tubes at the sam e time. The unit does not require recalibration if you are replacing the old sources with new sources of the same wavelength. 1. Purchase five tubes from UVP. 2. ...

  • UVPS 81-0112-01 Rev C - page 16

    UV CROSSLINKERS UVP 5. Cut all of the cable ties coming from the tube holder following the two wires from the tube holder to where it terminates. Note: ther e will be two diff erent termination points for each tube holder. Once this has been located, cut the wires (marking them to ensure easy replacement). 6. The tube holder(s) can now be removed b ...

  • UVPS 81-0112-01 Rev C - page 17

    UV CROSSLINKERS UVP 6. To secure the starter holder, place the holder in your right hand and with your left hand in the chamber, press the now loosened reflector adjacent to the holder inwards. This will push the black spigots inwards thereby securing the starter holder. 7. Re-fit the reflector. Replacement of fuse holders These are located at the ...

  • UVPS 81-0112-01 Rev C - page 18

    UV CROSSLINKERS UVP Replacement of UV sensor UVP recommends returning the Crosslinker to the UVP factory for service. This is located on the right hand side of the internal chassis assembly at the back near the base. 1. Unplug the sensor connector from the PCB board marked J2. 2. Release the sensor by rem oving the location screw with a screwdriver ...

  • UVPS 81-0112-01 Rev C - page 19

    UV CROSSLINKERS UVP C. Operations With The Internal Chassis Assembly Removed Removal of the assembly 1. Release the wire to the micro switch located on the bottom left internal chassis assem bly. 2. Release the main power cable to ballasts - black wire with connector (color red). 3. Release all eight nuts positioned f our on top of the unit, two se ...

  • UVPS 81-0112-01 Rev C - page 20

    UV CROSSLINKERS UVP E. General Items Changing fuses The fuses are located on the back panel and are marked with a self-adhesive sticker. 1. Turn fuse cap 15 degrees anti-clockwise. 2. Pull fuse and cap from the holder. 3. Pull fuse from the cap. 4. Replace the fuse. 5. Replace fuse cap and fuse into holder and turn 15 degrees clockwise. Changing pl ...

  • UVPS 81-0112-01 Rev C - page 21

    UV CROSSLINKERS UVP Schematic Wiring Diagram - 19 - ...

  • UVPS 81-0112-01 Rev C - page 22

    UV CROSSLINKERS UVP Internet: www.uvp.com Corporate Headquarters : UVP, Inc. 2066 W. 11th Street Upland, CA 91786 Tel: (800)452-6788 or (909)946-3197 Fax: (909)946-3597 / E-Mail: info@uvp.com European Operations: Ultra-Violet Products Ltd. Unit 1, Trinity Hall Estate, Nuffield Road Cambridge CB4 1TG United Kingdom Tel: +44(0)1223 420022 Fax: +44(0) ...

Fabbricante UVPS Categoria Light Therapy Device

I documenti che otteniamo dal fabbricante del dispositivoUVPS 81-0112-01 Rev C possono essere suddivisi in diversi gruppi. Questi sono tra gli altri:
- disegni tecnici UVPS
- manuali per l’uso 81-0112-01 Rev C
- schede prodotto UVPS
- opuscoli
- o etichette energetiche UVPS 81-0112-01 Rev C
Tutti sono importanti, ma le informazioni più importanti dal punto di vista di utilizzo del dispositivo possono essere trovate nel manuale per l’uso UVPS 81-0112-01 Rev C.

Un gruppo di documenti denominato manuali per l’uso, è anche suddiviso in tipi più specifici, come: Manuali di installazione UVPS 81-0112-01 Rev C, manuali di manutenzione, brevi manuali o manuali utente UVPS 81-0112-01 Rev C. A seconda delle esigenze, si dovrebbe cercare il documento necessario. Nel nostro sito web, puoi visualizzare il manuale più popolare per l'uso del prodotto UVPS 81-0112-01 Rev C.

Manuali per l’uso simili

Manuale completo del dispositivo UVPS 81-0112-01 Rev C, come deve essere?
Il manuale per l’uso, denominato anche istruzioni per l’uso, o semplicemente il manuale è un documento tecnico progettato per aiutare ad utilizzare UVPS 81-0112-01 Rev C dagli utenti. I manuali sono di solito scritti da uno scrittore tecnico, ma in un linguaggio accessibile a tutti gli utenti UVPS 81-0112-01 Rev C.

Il manuale per l’uso completo UVPS, dovrebbe includere vari elementi di base. Alcuni di loro sono meno importanti, come ad esempio: copertina / pagina del titolo o pagina di autore. Tuttavia, la parte rimanente, dovrebbe fornire informazioni importanti dal punto di vista dell'utente.

1. Introduzione e la guida su come utilizzare il manuale UVPS 81-0112-01 Rev C - All'inizio di ogni manuale, dovremmo trovare indizi su come utilizzare il documento. Dovrebbe contenere informazioni sulla posizione dell’indice UVPS 81-0112-01 Rev C, FAQ o i problemi più comuni : i punti che sono più spesso cercati dagli utenti di ogni manuale
2. Indice - elenco di tutti i suggerimenti per UVPS 81-0112-01 Rev C che possono essere trovati nel documento corrente
3. Suggerimenti su come utilizzare le funzioni di base del dispositivo UVPS 81-0112-01 Rev C - che ci dovrebbe aiutare nei primi passi durante l'uso UVPS 81-0112-01 Rev C
4. Troubleshooting - sequenza sistematica delle attività che vi aiuterà a diagnosticare e, successivamente, risolvere i problemi più importanti del UVPS 81-0112-01 Rev C
5. FAQ - domande più frequenti
6. Dettagli di contatto Informazioni su dove cercare il contatto con il fabbricante / servizio UVPS 81-0112-01 Rev C in un dato paese, se il problema non può essere risolto da noi stessi.

Hai una domanda su UVPS 81-0112-01 Rev C?

Utilizza il modulo sottostante

Se non hai risolto il problema con UVPS 81-0112-01 Rev C con l'aiuto del manuale che hai trovato, fai una domanda utilizzando il modulo sottostante. Se un utente ha avuto un problema simile con UVPS 81-0112-01 Rev C è probabile che voglia condividere il modo di risolverlo.

Riscrivi il testo dall’immagine

Commenti (0)