    Ultraviolet Crosslinkers Operating Instructions and Service Manual IMPORTANT: Please read these instructions before operating your UVP UV Crosslinker to familiarize yourself with its operation. 81-0112-01 Rev C ...

  • UVPS 81-0112-01 Rev C - page 2

    UV CROSSLINKERS UVP CONTENTS 1. Introduction 1 2. Important Safety Information 1 3. Description and Specifications 1 4. Operation 3 5. Applications 6 6. Maintenance, Care and Cleaning 7 7. Changing the UV Wavelength 8 8. Calibration Procedure 10 9. Replacement Parts 12 10. General Servicing Procedures 13 Schematic Diagram 19 ...

  • UVPS 81-0112-01 Rev C - page 3

    UV CROSSLINKERS UVP 1. INTRODUCTION UVP Crosslinkers offer researchers an instrume nt to quickly, safely and efficiently expose samples to a controlled amount of ultraviolet radiation. CL-1000 series features a lay-down type drawer while the CX-2000 has a pull-out dr awer. Exposure of samples to UV provides for the following: • Crosslinking of DN ...

  • UVPS 81-0112-01 Rev C - page 4

    UV CROSSLINKERS UVP Features • Microprocessor controlled/UV sensor feedback system • Multiple set functions- preset UV energy exposure preset UV time exposure user set UV energy exposure user set UV time exposure • Maximum UV energy exposure setting of 999,900 m icrojoules per cm2 • Maximum UV tim e exposure setting of 999.9 minutes • Int ...

  • UVPS 81-0112-01 Rev C - page 5

    UV CROSSLINKERS UVP 4. OPERATION 1. Place the crosslinker on a level work surface. Be sure there is enough room in front for the door to open easily. 2. Plug the female end of the power cord into the crosslinker. 3. Plug the male end of the power cord into a properly grounded electrical outlet. The proper operating voltage of the crosslinker is fou ...

  • UVPS 81-0112-01 Rev C - page 6

    UV CROSSLINKERS UVP The new setting will appear on the LED display. Push ENTER on the touchpad. The new setting is now installed. B. Preset ultraviolet time exposure Push PRESET and then TIME on the tactile touchpad. The red light at each position should now be lit and the prepr ogramm ed UV exposure time of 2 minutes should be displayed in the LED ...

  • UVPS 81-0112-01 Rev C - page 7

    UV CROSSLINKERS UVP Push START on the touchpad. After a slight delay to energize the UV tubes, The LED will begin to countdown. the unit will stop automatically at the end of the exposure cycle and will beep five times. Exposure is now complete. D. User set UV time exposure Setting your own UV time exposure can be achieved as follows: Push the TIME ...

  • UVPS 81-0112-01 Rev C - page 8

    UV CROSSLINKERS UVP 5. APPLICATIONS The Crosslinker is a multi-purpose ultraviolet e xposure instrum ent for use in the laboratory. A wide variety of applications for ultraviolet radiation exist in the laboratory. A. UV crosslinking of DNA and RNA by covale ntly binding nucleic acids to transfer membranes nitrocellulose, nylon or nylon-reinforced n ...

  • UVPS 81-0112-01 Rev C - page 9

    UV CROSSLINKERS UVP 6. MAINTENANCE, CARE AND CLEANING OF THE CROSSLINKER The UVP UV Crosslinker is built to provide you w ith trouble-free operation. To ensure correct service: NOTE: ALWAYS UNPLUG UNIT FROM ELEC TRICAL SUPPLY BEFORE CLEANING OR DRYING. 1. Wipe ANY water from inside and outside the unit with a soft cloth or sponge. 2. Use soap and w ...

  • UVPS 81-0112-01 Rev C - page 10

    UV CROSSLINKERS UVP 7. CHANGING THE UV WAVELENGTH A UV Crosslinker is purchased with shortwave, l ongwave or midrange tubes. However, if user requirements and applications change, the Crosslinkers provide you with the unique ability to change the ultraviolet wavelength and recalibrate the UV sensor and microprocessor to the new UV wavelength. This ...

  • UVPS 81-0112-01 Rev C - page 11

    UV CROSSLINKERS UVP 11. The previous calibrated operation of the crosslinker will continue following any unsuccessful recalibration. Replace the new UV wavelength tubes with the previously removed old wavelength tubes. 12. Successful recalibration to the NEW UV wavelength changes the values in the microprocessor and numeric setting value is display ...

  • UVPS 81-0112-01 Rev C - page 12

    UV CROSSLINKERS UVP 8. CROSSLINKER CALIBRATION PROCEDURE Equipment necessary: Chart recorder Two probes with banana plug adapters One radiometer with UVX-25 calibrated sensor Cable with banana plug and jack adaptor to connect radiometer to chart-recorder Power source (100V, 115V or 230V) Procedure: 1. Connect the crosslinker to the power source. Do ...

  • UVPS 81-0112-01 Rev C - page 13

    UV CROSSLINKERS UVP 9. Run the Crosslinker in calibration mode by depressing the STOP/RESET button and turning the power switch on at the same time. At this point, both pens should start recording information on paper. Som etimes either pen will record zero inform ation regardless of the lights in the crosslinker be ing on, if this is the case, che ...

  • UVPS 81-0112-01 Rev C - page 14

    UV CROSSLINKERS UVP 9. REPLACEMENT PARTS AND SERVICING PROCEDURES PART NO. A. OPERATIONS WITHOUT REMOVING THE TOP COVER 5 x 34-0007-01 Changing tubes B. OPERATIONS WITH TOP COVER REMOVED Replacement of tube holders 5 x 36-0003-01 Replacement of starters 5 x 53-0001-02 or 53-0132-01 Replacement of starter holders 5 x 36-0005-01 Replacement of fuse h ...

  • UVPS 81-0112-01 Rev C - page 15

    UV CROSSLINKERS UVP 10. GENERAL SERVICING PROCEDURES A. Operations Without Removing The Top Cover Changing the Tubes It is recommended to change all FIVE 8-W att 254nm UV tubes at the sam e time. The unit does not require recalibration if you are replacing the old sources with new sources of the same wavelength. 1. Purchase five tubes from UVP. 2. ...

  • UVPS 81-0112-01 Rev C - page 16

    UV CROSSLINKERS UVP 5. Cut all of the cable ties coming from the tube holder following the two wires from the tube holder to where it terminates. Note: ther e will be two diff erent termination points for each tube holder. Once this has been located, cut the wires (marking them to ensure easy replacement). 6. The tube holder(s) can now be removed b ...

  • UVPS 81-0112-01 Rev C - page 17

    UV CROSSLINKERS UVP 6. To secure the starter holder, place the holder in your right hand and with your left hand in the chamber, press the now loosened reflector adjacent to the holder inwards. This will push the black spigots inwards thereby securing the starter holder. 7. Re-fit the reflector. Replacement of fuse holders These are located at the ...

  • UVPS 81-0112-01 Rev C - page 18

    UV CROSSLINKERS UVP Replacement of UV sensor UVP recommends returning the Crosslinker to the UVP factory for service. This is located on the right hand side of the internal chassis assembly at the back near the base. 1. Unplug the sensor connector from the PCB board marked J2. 2. Release the sensor by rem oving the location screw with a screwdriver ...

  • UVPS 81-0112-01 Rev C - page 19

    UV CROSSLINKERS UVP C. Operations With The Internal Chassis Assembly Removed Removal of the assembly 1. Release the wire to the micro switch located on the bottom left internal chassis assem bly. 2. Release the main power cable to ballasts - black wire with connector (color red). 3. Release all eight nuts positioned f our on top of the unit, two se ...

  • UVPS 81-0112-01 Rev C - page 20

    UV CROSSLINKERS UVP E. General Items Changing fuses The fuses are located on the back panel and are marked with a self-adhesive sticker. 1. Turn fuse cap 15 degrees anti-clockwise. 2. Pull fuse and cap from the holder. 3. Pull fuse from the cap. 4. Replace the fuse. 5. Replace fuse cap and fuse into holder and turn 15 degrees clockwise. Changing pl ...

  • UVPS 81-0112-01 Rev C - page 21

    UV CROSSLINKERS UVP Schematic Wiring Diagram - 19 - ...

  • UVPS 81-0112-01 Rev C - page 22

    UV CROSSLINKERS UVP Internet: www.uvp.com Corporate Headquarters : UVP, Inc. 2066 W. 11th Street Upland, CA 91786 Tel: (800)452-6788 or (909)946-3197 Fax: (909)946-3597 / E-Mail: info@uvp.com European Operations: Ultra-Violet Products Ltd. Unit 1, Trinity Hall Estate, Nuffield Road Cambridge CB4 1TG United Kingdom Tel: +44(0)1223 420022 Fax: +44(0) ...

メーカー UVPS カテゴリー Light Therapy Device

UVPS 81-0112-01 Rev Cのメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- 81-0112-01 Rev Cの取扱説明書
- UVPSの製品カード
- パンフレット
- またはUVPS 81-0112-01 Rev Cの消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、UVPS 81-0112-01 Rev Cの取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、UVPS 81-0112-01 Rev Cの取り付け説明書、サービスマニュアル、簡易説明書、またはUVPS 81-0112-01 Rev Cのユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、UVPS 81-0112-01 Rev Cの製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

UVPS 81-0112-01 Rev Cデバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがUVPS 81-0112-01 Rev Cを使用するのを助ける技術的文書のことです。説明書は通常、全てのUVPS 81-0112-01 Rev Cユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

UVPSの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. UVPS 81-0112-01 Rev Cの説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははUVPS 81-0112-01 Rev Cの目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。UVPS 81-0112-01 Rev Cに関してこのドキュメントで見つけることができる全てのヒントの目次
3. UVPS 81-0112-01 Rev Cデバイスの基本機能を使うにあたってのヒント。 UVPS 81-0112-01 Rev Cのユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。UVPS 81-0112-01 Rev Cに関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるUVPS 81-0112-01 Rev Cのメーカー/サービスへの連絡先に関する情報。

UVPS 81-0112-01 Rev Cについてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもUVPS 81-0112-01 Rev Cの問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがUVPS 81-0112-01 Rev Cで同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)