    Ultraviolet Crosslinkers Operating Instructions and Service Manual IMPORTANT: Please read these instructions before operating your UVP UV Crosslinker to familiarize yourself with its operation. 81-0112-01 Rev C ...

  • UVPS 81-0112-01 Rev C - page 2

    UV CROSSLINKERS UVP CONTENTS 1. Introduction 1 2. Important Safety Information 1 3. Description and Specifications 1 4. Operation 3 5. Applications 6 6. Maintenance, Care and Cleaning 7 7. Changing the UV Wavelength 8 8. Calibration Procedure 10 9. Replacement Parts 12 10. General Servicing Procedures 13 Schematic Diagram 19 ...

  • UVPS 81-0112-01 Rev C - page 3

    UV CROSSLINKERS UVP 1. INTRODUCTION UVP Crosslinkers offer researchers an instrume nt to quickly, safely and efficiently expose samples to a controlled amount of ultraviolet radiation. CL-1000 series features a lay-down type drawer while the CX-2000 has a pull-out dr awer. Exposure of samples to UV provides for the following: • Crosslinking of DN ...

  • UVPS 81-0112-01 Rev C - page 4

    UV CROSSLINKERS UVP Features • Microprocessor controlled/UV sensor feedback system • Multiple set functions- preset UV energy exposure preset UV time exposure user set UV energy exposure user set UV time exposure • Maximum UV energy exposure setting of 999,900 m icrojoules per cm2 • Maximum UV tim e exposure setting of 999.9 minutes • Int ...

  • UVPS 81-0112-01 Rev C - page 5

    UV CROSSLINKERS UVP 4. OPERATION 1. Place the crosslinker on a level work surface. Be sure there is enough room in front for the door to open easily. 2. Plug the female end of the power cord into the crosslinker. 3. Plug the male end of the power cord into a properly grounded electrical outlet. The proper operating voltage of the crosslinker is fou ...

  • UVPS 81-0112-01 Rev C - page 6

    UV CROSSLINKERS UVP The new setting will appear on the LED display. Push ENTER on the touchpad. The new setting is now installed. B. Preset ultraviolet time exposure Push PRESET and then TIME on the tactile touchpad. The red light at each position should now be lit and the prepr ogramm ed UV exposure time of 2 minutes should be displayed in the LED ...

  • UVPS 81-0112-01 Rev C - page 7

    UV CROSSLINKERS UVP Push START on the touchpad. After a slight delay to energize the UV tubes, The LED will begin to countdown. the unit will stop automatically at the end of the exposure cycle and will beep five times. Exposure is now complete. D. User set UV time exposure Setting your own UV time exposure can be achieved as follows: Push the TIME ...

  • UVPS 81-0112-01 Rev C - page 8

    UV CROSSLINKERS UVP 5. APPLICATIONS The Crosslinker is a multi-purpose ultraviolet e xposure instrum ent for use in the laboratory. A wide variety of applications for ultraviolet radiation exist in the laboratory. A. UV crosslinking of DNA and RNA by covale ntly binding nucleic acids to transfer membranes nitrocellulose, nylon or nylon-reinforced n ...

  • UVPS 81-0112-01 Rev C - page 9

    UV CROSSLINKERS UVP 6. MAINTENANCE, CARE AND CLEANING OF THE CROSSLINKER The UVP UV Crosslinker is built to provide you w ith trouble-free operation. To ensure correct service: NOTE: ALWAYS UNPLUG UNIT FROM ELEC TRICAL SUPPLY BEFORE CLEANING OR DRYING. 1. Wipe ANY water from inside and outside the unit with a soft cloth or sponge. 2. Use soap and w ...

  • UVPS 81-0112-01 Rev C - page 10

    UV CROSSLINKERS UVP 7. CHANGING THE UV WAVELENGTH A UV Crosslinker is purchased with shortwave, l ongwave or midrange tubes. However, if user requirements and applications change, the Crosslinkers provide you with the unique ability to change the ultraviolet wavelength and recalibrate the UV sensor and microprocessor to the new UV wavelength. This ...

  • UVPS 81-0112-01 Rev C - page 11

    UV CROSSLINKERS UVP 11. The previous calibrated operation of the crosslinker will continue following any unsuccessful recalibration. Replace the new UV wavelength tubes with the previously removed old wavelength tubes. 12. Successful recalibration to the NEW UV wavelength changes the values in the microprocessor and numeric setting value is display ...

  • UVPS 81-0112-01 Rev C - page 12

    UV CROSSLINKERS UVP 8. CROSSLINKER CALIBRATION PROCEDURE Equipment necessary: Chart recorder Two probes with banana plug adapters One radiometer with UVX-25 calibrated sensor Cable with banana plug and jack adaptor to connect radiometer to chart-recorder Power source (100V, 115V or 230V) Procedure: 1. Connect the crosslinker to the power source. Do ...

  • UVPS 81-0112-01 Rev C - page 13

    UV CROSSLINKERS UVP 9. Run the Crosslinker in calibration mode by depressing the STOP/RESET button and turning the power switch on at the same time. At this point, both pens should start recording information on paper. Som etimes either pen will record zero inform ation regardless of the lights in the crosslinker be ing on, if this is the case, che ...

  • UVPS 81-0112-01 Rev C - page 14

    UV CROSSLINKERS UVP 9. REPLACEMENT PARTS AND SERVICING PROCEDURES PART NO. A. OPERATIONS WITHOUT REMOVING THE TOP COVER 5 x 34-0007-01 Changing tubes B. OPERATIONS WITH TOP COVER REMOVED Replacement of tube holders 5 x 36-0003-01 Replacement of starters 5 x 53-0001-02 or 53-0132-01 Replacement of starter holders 5 x 36-0005-01 Replacement of fuse h ...

  • UVPS 81-0112-01 Rev C - page 15

    UV CROSSLINKERS UVP 10. GENERAL SERVICING PROCEDURES A. Operations Without Removing The Top Cover Changing the Tubes It is recommended to change all FIVE 8-W att 254nm UV tubes at the sam e time. The unit does not require recalibration if you are replacing the old sources with new sources of the same wavelength. 1. Purchase five tubes from UVP. 2. ...

  • UVPS 81-0112-01 Rev C - page 16

    UV CROSSLINKERS UVP 5. Cut all of the cable ties coming from the tube holder following the two wires from the tube holder to where it terminates. Note: ther e will be two diff erent termination points for each tube holder. Once this has been located, cut the wires (marking them to ensure easy replacement). 6. The tube holder(s) can now be removed b ...

  • UVPS 81-0112-01 Rev C - page 17

    UV CROSSLINKERS UVP 6. To secure the starter holder, place the holder in your right hand and with your left hand in the chamber, press the now loosened reflector adjacent to the holder inwards. This will push the black spigots inwards thereby securing the starter holder. 7. Re-fit the reflector. Replacement of fuse holders These are located at the ...

  • UVPS 81-0112-01 Rev C - page 18

    UV CROSSLINKERS UVP Replacement of UV sensor UVP recommends returning the Crosslinker to the UVP factory for service. This is located on the right hand side of the internal chassis assembly at the back near the base. 1. Unplug the sensor connector from the PCB board marked J2. 2. Release the sensor by rem oving the location screw with a screwdriver ...

  • UVPS 81-0112-01 Rev C - page 19

    UV CROSSLINKERS UVP C. Operations With The Internal Chassis Assembly Removed Removal of the assembly 1. Release the wire to the micro switch located on the bottom left internal chassis assem bly. 2. Release the main power cable to ballasts - black wire with connector (color red). 3. Release all eight nuts positioned f our on top of the unit, two se ...

  • UVPS 81-0112-01 Rev C - page 20

    UV CROSSLINKERS UVP E. General Items Changing fuses The fuses are located on the back panel and are marked with a self-adhesive sticker. 1. Turn fuse cap 15 degrees anti-clockwise. 2. Pull fuse and cap from the holder. 3. Pull fuse from the cap. 4. Replace the fuse. 5. Replace fuse cap and fuse into holder and turn 15 degrees clockwise. Changing pl ...

  • UVPS 81-0112-01 Rev C - page 21

    UV CROSSLINKERS UVP Schematic Wiring Diagram - 19 - ...

  • UVPS 81-0112-01 Rev C - page 22

    UV CROSSLINKERS UVP Internet: www.uvp.com Corporate Headquarters : UVP, Inc. 2066 W. 11th Street Upland, CA 91786 Tel: (800)452-6788 or (909)946-3197 Fax: (909)946-3597 / E-Mail: info@uvp.com European Operations: Ultra-Violet Products Ltd. Unit 1, Trinity Hall Estate, Nuffield Road Cambridge CB4 1TG United Kingdom Tel: +44(0)1223 420022 Fax: +44(0) ...

Manufacturer UVPS Category Light Therapy Device

Documents that we receive from a manufacturer of a UVPS 81-0112-01 Rev C can be divided into several groups. They are, among others:
- UVPS technical drawings
- 81-0112-01 Rev C manuals
- UVPS product data sheets
- information booklets
- or energy labels UVPS 81-0112-01 Rev C
All of them are important, but the most important information from the point of view of use of the device are in the user manual UVPS 81-0112-01 Rev C.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals UVPS 81-0112-01 Rev C, service manual, brief instructions and user manuals UVPS 81-0112-01 Rev C. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product UVPS 81-0112-01 Rev C.

Similar manuals

A complete manual for the device UVPS 81-0112-01 Rev C, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use UVPS 81-0112-01 Rev C by users. Manuals are usually written by a technical writer, but in a language understandable to all users of UVPS 81-0112-01 Rev C.

A complete UVPS manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual UVPS 81-0112-01 Rev C - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the UVPS 81-0112-01 Rev C, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the UVPS 81-0112-01 Rev C, that we can find in the current document
3. Tips how to use the basic functions of the device UVPS 81-0112-01 Rev C - which should help us in our first steps of using UVPS 81-0112-01 Rev C
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with UVPS 81-0112-01 Rev C
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of UVPS 81-0112-01 Rev C in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning UVPS 81-0112-01 Rev C?

Use the form below

If you did not solve your problem by using a manual UVPS 81-0112-01 Rev C, ask a question using the form below. If a user had a similar problem with UVPS 81-0112-01 Rev C it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)