Airflow Systems IAQ10の取扱説明書

20ページ 0.18 mb
ダウンロード

ページに移動 of 20

Summary
  • Airflow Systems IAQ10 - page 1

    CO 2 Meter Model IAQ10 OWNER’S MANUAL ...

  • Airflow Systems IAQ10 - page 2

    ii LIMITATION OF WA RRANTY A ND LIABILITY Seller warrants the goods sold hereunder, under normal use and ser vice as described in the operator's manual, s hall be free from defects in work manship and m aterial for twenty-four (24) months, or the length of tim e specif ied in the operator's manual, f rom the date of shipment to the c usto ...

  • Airflow Systems IAQ10 - page 3

    1 Table of Contents LIMITATION OF WARRANTY AN D LIABILITY ...... ii SECTION 1 General Descri ption ..................................... 2 SECTION 2 Theory of Opera tion: NDIR Sensor ........... 2 SECTION 3 Safety .............................................................. 3 SECTION 4 Setting-Up the Model IAQ10 ...................... 4 Supplying ...

  • Airflow Systems IAQ10 - page 4

    2 SECTION 1 General Description The Model IAQ10 is a hand-held meter to measure Carbon Dioxide (CO 2 ), displayed in units of parts per million (ppm). An NDIR sensor in the top end of the meter measures gas content by diffusion through sensing holes. The IAQ10 ships in a pouch that prov ides a small amount of protection to the display and the sensi ...

  • Airflow Systems IAQ10 - page 5

    3 concentrations. This technique is called non-dispersive infrared (NDIR) detection. SECTION 3 Safety When using the IAQ10 to check for CO 2 values, make certain that you can safely raise and hold the instrument while making measurements. Be especially careful when working on a ladder. Observe all necessary precautions so that the unit does not bec ...

  • Airflow Systems IAQ10 - page 6

    4 SECTION 4 Setting-Up the Model IAQ10 Supplying Power The IAQ10 can be powered in one of two ways: four AA-size batteries or the optiona l AC adapter. Installing the Batteries Insert four AA batteries as indicated by the diagram located on the inside of the battery compartment. The Model IAQ10 is de- signed to operate only with alkaline batteries. ...

  • Airflow Systems IAQ10 - page 7

    5 SECTION 5 Operation of the Model IAQ10 Note : Do not hold the instrument close to your face. Humans exhale CO 2 and this will affect the accuracy of the readings. When pressing the keys on the keypad, the Model IAQ10 will beep to confirm the function. If you press a key and the Model IAQ10 does not beep, the Model IAQ10 does not allow that functi ...

  • Airflow Systems IAQ10 - page 8

    6 calibration. When the Model IAQ10 completes the internal self-check, it will display the approximate percentage of battery life remaining. Note : The sensor needs up to 10 minutes to stabilize in still air before displayed readings can be considered accurate. Moving the instrument may decrease this stabilization time. READ Key Carbon dioxide conc ...

  • Airflow Systems IAQ10 - page 9

    7 The display updates every second and each reading is the average of five samples. ( 5 ) Key The instrument only displays CO 2 . RECALL ( 6 ) Key Press the RECALL key repeatedly to display: minimum value ( MIN ), maximum value ( MAX ), average value ( AVG ), and number of seconds sampled ( COUNT ). Press the READ key to return to normal measuring ...

  • Airflow Systems IAQ10 - page 10

    8 SECTION 6 Maintenance of the Model IAQ10 The Model IAQ10 requires very little maintenance to keep it performing well. Case If the instrument case or storage case needs cleaning, wipe it off with a soft, damp cloth. Never submerge the IAQ10 in any liquid or allow any liquid to enter the sensing holes. Storage When storing the Model IAQ10 for more ...

  • Airflow Systems IAQ10 - page 11

    9 Before sending your instrument for calibration or repair, you should call Customer Service. The service department will provide you with the cost of service or calibration, Return Materia l Authorization (RMA) number, and shipping instructions. Please have the following information available when you call: • Owner’s name, address, and phone n ...

  • Airflow Systems IAQ10 - page 12

    10 you with a NIST (US Na tional Institute of Standards and Technology) traceable certificate. This “annual checkup” assures you of consistently accurate readings; it is especially important in applications where strict calibration records must be main- tained. Send the instrument to TSI Instruments Ltd. prepaid. Securely package your instrumen ...

  • Airflow Systems IAQ10 - page 13

    11 1. Turn the instrument off. Under the batteries, move DIP switch # 7 to ON. Turn the instrument back on. 2. Press and hold the READ key to begin the calibration procedure. The display will begin a countdown from 5 to 0. Release the key when the display reads 0. If the key is released too soon or too late, the instrument will go back to sam- plin ...

  • Airflow Systems IAQ10 - page 14

    12 indicates “ SPAN ” and the span concen- tration. 6. Install the regulator on the span calibra- tion cylinder and connect tubing from the regulator to the inlet fitting on the top of the Model IAQ10. 7. Use 5 and 6 to adjust the concentra- tion displayed on the Model IAQ10 to match the concentration on the span gas cylinder. 8. Press and rele ...

  • Airflow Systems IAQ10 - page 15

    13 SECTION 8 Troubleshooting of the Model IAQ10 This table lists the symptoms, possibl e causes, and recommended solutions for the Model IAQ10. Symptom Possible Causes Corrective A ction(s) No display Unit not switched on Switch on the unit Low or dead batteries Replace the batteries Dirty battery contacts Clean the battery contacts Batteries insta ...

  • Airflow Systems IAQ10 - page 16

    14 Symptom Possible Causes Corrective A ction(s) Display reads “LO” Batteries are low Replace the batteries Note: If your symptoms are not remedied by the suggested corrective action, call TSI Instruments Ltd. SECTION 9 Internal DIP Sw itch Settings To access the DIP switches, remove the batteries from the battery compartment. On the inside of ...

  • Airflow Systems IAQ10 - page 17

    15 DIP Sw itch OFF ON 1-3 Must be OFF ----- 4 not assigned not assigned 5 not assigned not assigned 6 Beep is turned OFF Beep is turned ON 7 Normal mode Field Verifica- tion mode 8 not assigned not assigned - The ON position is away from the batteries and OFF is towards the batteries. - Switch 1 is towards the top of the case and Switch 8 is neares ...

  • Airflow Systems IAQ10 - page 18

    ...

  • Airflow Systems IAQ10 - page 19

    Specifications (subject to change without n otice) CO 2 Sensor type: Dual wavelength detector with non - dispersive infrared (NDIR) sensor Range: 0 to 5000 ppm Accuracy ±3% of reading or ±50 ppm, which ever is greater 1, 2 Resolution 1 ppm Response time <10 minutes in still air Operating Temperat ure Range: 5 to 70°C External Meter Dimensions ...

  • Airflow Systems IAQ10 - page 20

    TSI Instruments Ltd. October 2005 Lancaster Road Printed in USA Cressex Business Park Part No. 1980532, Rev. A High Wycombe Buckinghamshire Tel: +44 (0) 149 4 459200 HP12 3QP England Fax: +44 (0) 149 4 459700 E-mail: info@airflowinstruments.co.uk Copyright © 2005 Web: www.a irf lo wins tr ume nts .c o. uk TSI Incorporated ...

メーカー Airflow Systems カテゴリー Carbon Monoxide Alarm

Airflow Systems IAQ10のメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- IAQ10の取扱説明書
- Airflow Systemsの製品カード
- パンフレット
- またはAirflow Systems IAQ10の消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Airflow Systems IAQ10の取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Airflow Systems IAQ10の取り付け説明書、サービスマニュアル、簡易説明書、またはAirflow Systems IAQ10のユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Airflow Systems IAQ10の製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Airflow Systems IAQ10デバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがAirflow Systems IAQ10を使用するのを助ける技術的文書のことです。説明書は通常、全てのAirflow Systems IAQ10ユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Airflow Systemsの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Airflow Systems IAQ10の説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははAirflow Systems IAQ10の目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Airflow Systems IAQ10に関してこのドキュメントで見つけることができる全てのヒントの目次
3. Airflow Systems IAQ10デバイスの基本機能を使うにあたってのヒント。 Airflow Systems IAQ10のユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Airflow Systems IAQ10に関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるAirflow Systems IAQ10のメーカー/サービスへの連絡先に関する情報。

Airflow Systems IAQ10についてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもAirflow Systems IAQ10の問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがAirflow Systems IAQ10で同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)