    Shop online at FMG-3000 S ERIES Blind version Magmeter U ser ’ s Guide ome ga.com om e g a .c om TM ® www .omega.com e-mail: http://omegamanual.info For latest information and product manual visit www .omegamanual.info. ISO 9001 CERTIFIED CORPORA TE QUALITY ST AMFORD , CT ISO 9002 CERTIFIED CORPORA TE QUALITY MANCHESTER, UK ...

  • Omega FMG3000 - page 2

    It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certifi cation of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certifi cation. The information contained in this document is believed to be correct, but ...

  • Omega FMG3000 - page 3

    1 1 2 3 4 JP2 FREQUENCY OUT DO NOT REMOVE Connect output signals and power to this 4-terminal block. 2. 4. 3. 3. Sensor T ype Pipe Size Jumper Position ½ in. to 2½ in. FMG3001 DN15 to DN65 FMG3002 3 in. to 4 in. DN80 to DN100 FMG3101 5 in. to 6 in. FMG3102 DN125 to DN150 8 in. DN200 FMG3201 10 in. to 12 in. FMG3202 DN250 to DN300 1. Omega FMG-300 ...

  • Omega FMG3000 - page 4

    2 SAFETY INSTRUCTIONS 1. Depressurize and vent system prior to installation or removal. 2. Confi rm chemical compatibility before use. 3. Do not exceed maximum temperature/pressure specifi cations. 4. Wear safety goggles or face shield during installation/service. 5. Do not alter product construction. 6. Disconnect power before attempting any ser ...

  • Omega FMG3000 - page 5

    3 4. Selecting a Location • The FMG-3000 requires a full pipe and a fully developed turbulent fl ow profi le for accurate measurement. • If the piping system harbors air pockets or bubbles, take steps to locate the sensor so the air pockets will not contact the electrodes. • In vertical installations, assemble the unit so the conduit ports ...

  • Omega FMG3000 - page 6

    4 Important: • The directional arrow on the body of the sensor must be pointed DOWNSTREAM. • The FLOW arrow decal can be placed directly on the pipe to identify the direction of fl ow . • Use a cable gland or a liquid tight connector to seal the cable ports from water intrusion. • The yellow housing may be reversed to align the conduit por ...

  • Omega FMG3000 - page 7

    5 6. General Installation and Grounding Tips Sensor conditioning The Magmeter output signal may be unstable immediately after installation. Allowing the sensor to soak in a full pipe (or in a ny container of water) for 24 hours will stabilize the performance. • V ery low conductivity fl uids may require a longer conditioning period. (The Magmete ...

  • Omega FMG3000 - page 8

    6 • The FMG-3001 and FMG-3101 Magmeters output an open collector frequency signal. • A 10 K Ω pull up resistor must be connected between terminals 1 and 2 if the magmeter is used with third-party equipment.. 8. Wiring the FMG-3001 and FMG-3101 with Frequency output 1 2 3 4 5-24 VDC Ground Not used Frequency Out Magmeter 10K Ω Install a pull ...

  • Omega FMG3000 - page 9

    7 9. Calibration Data: K-factors* and Full Scale Current V alues Plastic Installation Fittings: PVC T ees and Saddles PIPE SIZE FITTING K-Factor K-Factor 20 mA= 20 mA= (IN.) TYPE Gallons Liters* in GPM in LPM SCH 80 PVC TEES FOR SCH 80 PVC PIPE ½ FP-5305 2289.37 604.85 13.10 49.60 ¾ FP-5307 1430.41 377.92 20.97 79.38 1 FP-5310 876.86 231.67 34.21 ...

  • Omega FMG3000 - page 10

    8 PIPE SIZE FITTING K-Factor K-Factor 20 mA= 20 mA= (IN.) TYPE Gallons Liters* in GPM in LPM POL YPROPYLENE FITTINGS (DIN/ISO ,BS, ANSI) ½ FP-5105PO 2192.73 579.32 13.68 51.78 ¾ FP-5107PO 1327.81 350.81 22.59 85.52 1 FP-51 10PO 737.16 194.76 40.70 154.04 1¼ FP-51 12PO 453.46 1 19.81 66.16 250.41 1½ FP-51 15PO 275.03 72.66 109.08 412.86 2 FP-512 ...

  • Omega FMG3000 - page 11

    9 PIPE SIZE FITTING K-Factor K-Factor 20 mA= 20 mA= (IN.) TYPE Gallons Liters* in GPM in LPM CARBON STEEL TEES ON SCH 40 PIPE ½ FMG-5305CS 1572.66 415.50 19.08 72.20 ¾ FMG-5307CS 1086.73 287.1 1 27.61 104.49 1 FMG-5310CS 582.34 153.86 51.52 194.99 1¼ FMG-5312CS 377.48 99.73 79.48 300.81 1½ FMG-5315CS 267.79 70.75 1 12.03 424.02 2 FMG-5320CS 167 ...

  • Omega FMG3000 - page 12

    10 PIPE SIZE FITTING K-Factor K-Factor 20 mA= 20 mA= (IN.) TYPE Gallons Liters* in GPM in LPM BRONZE TEES ON SCH 40 PIPE 1 FP-5310BR 582.34 153.86 51.52 194.99 1¼ FP-5312BR 330.54 87.33 90.76 343.53 1½ FP-5315BR 254.76 67.31 1 17.76 445.71 2 FP-5320BR 157.36 41.58 190.64 721.58 COPPER TEES FITTING ON COPPER PIPE SCH K ½ FP-5305CU 2459.19 649.72 ...

  • Omega FMG3000 - page 13

    11 • Conductivity is less than 20 µS/cm. • Electronic component failure. • 4-20 mA is not scaled same as Loop device. • Loop device is not scaled same as Magmeter . • Range Jumper not placed correctly . • Defective Magmeter • Magmeter installed too close to upstream obstruction. • Magmeter electrodes are coated with solids. • Mag ...

  • Omega FMG3000 - page 14

    12 1 1. Ordering Information OMEGAMAG with Frequency Output FMG3001-PP OMEGAMAG, Blind, Freq, PP/316SS, ½-4 FMG3101-PP OMEGAMAG, Blind, Freq, PP/316SS, 5-8 FMG3201-PP OMEGAMAG, Blind, Freq, PP/316SS,10-12 FMG-3001-PVDF-HA OMEGAMAG, Blind, Freq, PVDF/Hast, ½-4 FMG3101-PVDF-HA OMEGAMAG, Blind, Freq, PVDF/Hast, 5-8 FMG3201-PVDF-HA OMEGAMAG, Blind, F ...

  • Omega FMG3000 - page 15

    W ARRANTY/DISCLAIMER OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA ’s W ARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA ’s customers ...

  • Omega FMG3000 - page 16

    6-2551.090-OM (C-4/07) M-4210/0407 Where Do I Find Everything I Need for Process Measurement and Control? OMEGA…Of Course! Shop online at www .omega.com TEMPERATURE • Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies • Wire: Thermocouple, RTD & Thermistor • Calibrators & Ice Point References • Recorder ...

メーカー Omega カテゴリー Camera Accessories

Omega FMG3000のメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- FMG3000の取扱説明書
- Omegaの製品カード
- パンフレット
- またはOmega FMG3000の消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Omega FMG3000の取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Omega FMG3000の取り付け説明書、サービスマニュアル、簡易説明書、またはOmega FMG3000のユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Omega FMG3000の製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Omega FMG3000デバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがOmega FMG3000を使用するのを助ける技術的文書のことです。説明書は通常、全てのOmega FMG3000ユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Omegaの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Omega FMG3000の説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははOmega FMG3000の目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Omega FMG3000に関してこのドキュメントで見つけることができる全てのヒントの目次
3. Omega FMG3000デバイスの基本機能を使うにあたってのヒント。 Omega FMG3000のユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Omega FMG3000に関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるOmega FMG3000のメーカー/サービスへの連絡先に関する情報。

Omega FMG3000についてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもOmega FMG3000の問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがOmega FMG3000で同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)