    OPERATION PROCEDURES FOR PHILIPS/FEI CM-20 TRANSMISSION ELECTRON MICROSCOPE General Comments Normally the m icroscope will be left in the power ON (standby ready to operate) condition; i.e., all circuits will be con tinuously energized and the microscope will be under full vacuum (MAIN, VACUUM, HIGH TENSION lamps on th e right control panel are all ...

  • Philips CM-20 - page 2

    (2) Hold the specimen holder with the airlock-op ening pin up. Carefully insert the specimen holder into the airlock entry and slide it in un til stop. At this point , pre-pumping of the airlock commences as indicated by illum ina tion of the red airlock indicator light. (3) After hearing a click sound, gr ape the specimen holder and turn clockwise ...

  • Philips CM-20 - page 3

    (4) Make sure the camera cover is correctly pl aced, and then press CAM AIR again to pump the camera chamber. You should be able to hear the mechanical pum p starting. Pull on the camera lid to ensure the vacuum is pulling. (5) Press READY, then TEM CAMERA, and CAM INIT. (6) Press RESET to reset film numb er (to 36), and press READY. (7) Press MODE ...

  • Philips CM-20 - page 4

    ROUTINE ALIGNMENT PROCEDUR ES FOR PHILIPS CM-20 TEM The following protocol is not intended to be spec ific. It is intended to be used as a mean to refresh one’s memory in carry out step by st ep procedures which are necessary to obtain reasonably good results with the transmission electron microscopes. More specific information can be obtained fr ...

  • Philips CM-20 - page 5

    (6) Set voltage at 20. Push HIGH TENSI ON button (the green LED should light) Slowly increase voltage from 20 kV to 160 kV using soft keys (wait 1 minute between each step). (7) After voltage reaches 160 kV, press FR EE CONTROL soft key. Use focus knob, slowly increase voltage from 160kV to 200kV. If it arcs (loud popping sound), use focus step siz ...

  • Philips CM-20 - page 6

    by first making crossover image with INTENSITY knob, then open up condenser II (INTENSITY)/expand the beam and close condenser II on the opposite side of crossover. If the beam appears to sw ing this indicates th e aperture is not properly centered. Center the aperture by moving the X and Y controls on the aperture housing. (5) Repeat the last step ...

  • Philips CM-20 - page 7

    (4) Condenser lens astigmatism correction - Press STIG to activate stigmator control. - Press COND soft key so that it is highlighted (the default is OBJ). - Condense the beam. - Use MULTIFUNCTION knobs to correct the astigmatism. The beam should be round when changing the setting of the INTENSI TY knob, or the filament image will be sharpest if fi ...

  • Philips CM-20 - page 8

    - Press ALGN to de-activate. (3) Voltage centering - Place a more or less symmetrical im age to the center. - Condense the beam at the center of screen, and spread the beam . - Press ROT CENTER and VOLT soft keys in ALGN. - Make sure it is in the VOLT m ode (check bottom right soft key). - Determine the voltage center from the m ovement of the enti ...

  • Philips CM-20 - page 9

    (5) Change camera length (by changi ng MAGNIFICATION knob) to 44.0 mm. (6) Realign the diffraction pattern. (7) Repeat steps 4 to 6 until the transm itted spot remains at (near) the center w hile changing camera length between 44.0 mm to 3.2 m . (8) Deactivate ALGN, and depress D to get out of diffraction mode. (iv) Insertion and centering objectiv ...

  • Philips CM-20 - page 10

    (2) Set a magnification at least 3 or 4 times hi gher than that will used for photographing. (3) Obtain the thinnest over-focus (Fresnel) fringe that can be clearly seen in the binocular, with beam spread. Caution: make sure you focus the binocular on the viewing screen. (4) Adjust the OJB STIG X and Y (MULTIFUNCTION keys) until the fringe fo llows ...

  • Philips CM-20 - page 11

    WAC 09/28/02 VI. COMPUTER ALIGNMENT AND AS TGMATISM CORRECTION FOR 200 kV 1. Preses PARAMETERS on TEM br ight field page 3 times 2. Press REMOTE CONTROL to highlight it (This w ill allow the com puter to talk to the TEM). 3. Turn on the PC monitor (on/o ff switch on back of m onitor). 4. If computer is already in execution mode, go to step 6 5. At ...

メーカー Philips カテゴリー Microscope & Magnifier

Philips CM-20のメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- CM-20の取扱説明書
- Philipsの製品カード
- パンフレット
- またはPhilips CM-20の消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Philips CM-20の取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Philips CM-20の取り付け説明書、サービスマニュアル、簡易説明書、またはPhilips CM-20のユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Philips CM-20の製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Philips CM-20デバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがPhilips CM-20を使用するのを助ける技術的文書のことです。説明書は通常、全てのPhilips CM-20ユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Philipsの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Philips CM-20の説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははPhilips CM-20の目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Philips CM-20に関してこのドキュメントで見つけることができる全てのヒントの目次
3. Philips CM-20デバイスの基本機能を使うにあたってのヒント。 Philips CM-20のユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Philips CM-20に関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるPhilips CM-20のメーカー/サービスへの連絡先に関する情報。

Philips CM-20についてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもPhilips CM-20の問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがPhilips CM-20で同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)